Asymmetric Outer Bow Length and Cervical Headgear Force System: 3D Analysis Using Finite Element Method.

Geramy A, Hassanpour M, Emadian Razavi ES - J Dent (Tehran) (2015)

Bottom Line: A 2.5 N force in horizontal plane was applied and the loading manner of each side of the outer bow was calculated trigonometrically using data from a volunteer.The 15mm difference in outer bow length caused the greatest difference in lateral (=0.21 N) and distal (= 1.008 N) forces and also generated moments (5.044 N.mm).A difference range of 1mm to 15 mm of length in cervical headgear can be considered as a safe length of outer bow shortening in clinical use.

View Article: PubMed Central - PubMed

Affiliation: Dental Research Center, Dentistry Research Institute, Tehran University of Medical Sciences, Tehran, Iran ; Professor, Department of Orthodontics, Tehran University of Medical Sciences, Tehran, Iran.

ABSTRACT

Objectives: This study sought to assess distal and lateral forces and moments of asymmetric headgears by variable outer bow lengths.

Materials and methods: Four 3D finite element method (FEM) models of a cervical headgear attached to the maxillary first molars were designed in SolidWorks 2010 software and transferred to ANSYS Workbench ver. 11 software. Models contained the first molars, their periodontal ligament (PDL), cancellous and cortical bones, a mesiodistal slice of the maxillae and the headgear. Models were the same except for the outer bow length in headgears. The headgear was symmetric in model 1. In models 2 to 4, the headgears were asymmetric in length with differences of 5mm, 10mm and 15mm, respectively. A 2.5 N force in horizontal plane was applied and the loading manner of each side of the outer bow was calculated trigonometrically using data from a volunteer.

Results: The 15mm difference in outer bow length caused the greatest difference in lateral (=0.21 N) and distal (= 1.008 N) forces and also generated moments (5.044 N.mm).

Conclusion: As the difference in outer bow length became greater, asymmetric effects increased. Greater distal force in the longer arm side was associated with greater lateral force towards the shorter arm side and more net yawing moment.

Clinical relevance: A difference range of 1mm to 15 mm of length in cervical headgear can be considered as a safe length of outer bow shortening in clinical use.

No MeSH data available.


Related in: MedlinePlus

The influence of asymmetric outer bow on distal force
© Copyright Policy
Related In: Results  -  Collection


getmorefigures.php?uid=PMC4663312&req=5

Figure 5: The influence of asymmetric outer bow on distal force

Mentions: Distal force had the same trend as the lateral force. But, It should be noticed that force differences were greater for distal forces than for lateral forces in the same length differences. It means that greater asymmetric effect was ob served due to distal than lateral force. Table 3 shows distal force and the effect produced by each asymmetric headgear (Figure 5).


Asymmetric Outer Bow Length and Cervical Headgear Force System: 3D Analysis Using Finite Element Method.

Geramy A, Hassanpour M, Emadian Razavi ES - J Dent (Tehran) (2015)

The influence of asymmetric outer bow on distal force
© Copyright Policy
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC4663312&req=5

Figure 5: The influence of asymmetric outer bow on distal force
Mentions: Distal force had the same trend as the lateral force. But, It should be noticed that force differences were greater for distal forces than for lateral forces in the same length differences. It means that greater asymmetric effect was ob served due to distal than lateral force. Table 3 shows distal force and the effect produced by each asymmetric headgear (Figure 5).

Bottom Line: A 2.5 N force in horizontal plane was applied and the loading manner of each side of the outer bow was calculated trigonometrically using data from a volunteer.The 15mm difference in outer bow length caused the greatest difference in lateral (=0.21 N) and distal (= 1.008 N) forces and also generated moments (5.044 N.mm).A difference range of 1mm to 15 mm of length in cervical headgear can be considered as a safe length of outer bow shortening in clinical use.

View Article: PubMed Central - PubMed

Affiliation: Dental Research Center, Dentistry Research Institute, Tehran University of Medical Sciences, Tehran, Iran ; Professor, Department of Orthodontics, Tehran University of Medical Sciences, Tehran, Iran.

ABSTRACT

Objectives: This study sought to assess distal and lateral forces and moments of asymmetric headgears by variable outer bow lengths.

Materials and methods: Four 3D finite element method (FEM) models of a cervical headgear attached to the maxillary first molars were designed in SolidWorks 2010 software and transferred to ANSYS Workbench ver. 11 software. Models contained the first molars, their periodontal ligament (PDL), cancellous and cortical bones, a mesiodistal slice of the maxillae and the headgear. Models were the same except for the outer bow length in headgears. The headgear was symmetric in model 1. In models 2 to 4, the headgears were asymmetric in length with differences of 5mm, 10mm and 15mm, respectively. A 2.5 N force in horizontal plane was applied and the loading manner of each side of the outer bow was calculated trigonometrically using data from a volunteer.

Results: The 15mm difference in outer bow length caused the greatest difference in lateral (=0.21 N) and distal (= 1.008 N) forces and also generated moments (5.044 N.mm).

Conclusion: As the difference in outer bow length became greater, asymmetric effects increased. Greater distal force in the longer arm side was associated with greater lateral force towards the shorter arm side and more net yawing moment.

Clinical relevance: A difference range of 1mm to 15 mm of length in cervical headgear can be considered as a safe length of outer bow shortening in clinical use.