Limits...
First Steps Toward Ultrasound-Based Motion Compensation for Imaging and Therapy: Calibration with an Optical System and 4D PET Imaging.

Schwaab J, Kurz C, Sarti C, Bongers A, Schoenahl F, Bert C, Debus J, Parodi K, Jenne JW - Front Oncol (2015)

Bottom Line: Furthermore, it is demonstrated that the US probe being within the PET field of view generally has no relevant influence on the image quality.The accuracy and precision of all the steps in the calibration workflow for US tracking-based 4D PET imaging are found to be in an acceptable range for clinical implementation.Eventually, we show in vitro that an US-based motion tracking in absolute room coordinates with a moving US transducer is feasible.

View Article: PubMed Central - PubMed

Affiliation: Mediri GmbH , Heidelberg , Germany.

ABSTRACT
Target motion, particularly in the abdomen, due to respiration or patient movement is still a challenge in many diagnostic and therapeutic processes. Hence, methods to detect and compensate this motion are required. Diagnostic ultrasound (US) represents a non-invasive and dose-free alternative to fluoroscopy, providing more information about internal target motion than respiration belt or optical tracking. The goal of this project is to develop an US-based motion tracking for real-time motion correction in radiation therapy and diagnostic imaging, notably in 4D positron emission tomography (PET). In this work, a workflow is established to enable the transformation of US tracking data to the coordinates of the treatment delivery or imaging system - even if the US probe is moving due to respiration. It is shown that the US tracking signal is equally adequate for 4D PET image reconstruction as the clinically used respiration belt and provides additional opportunities in this concern. Furthermore, it is demonstrated that the US probe being within the PET field of view generally has no relevant influence on the image quality. The accuracy and precision of all the steps in the calibration workflow for US tracking-based 4D PET imaging are found to be in an acceptable range for clinical implementation. Eventually, we show in vitro that an US-based motion tracking in absolute room coordinates with a moving US transducer is feasible.

No MeSH data available.


Related in: MedlinePlus

Experimental setup for investigation of the US transducer’s influence on PET image reconstruction. The three radioactive point sources were aligned in a horizontal diagonal line within the acrylic glass table. The measurement was repeated with the transducer in four different positions: (A) next to table (edge), (B) next to table (corner), (C) under table, and (D) on table.
© Copyright Policy
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC4663279&req=5

Figure 4: Experimental setup for investigation of the US transducer’s influence on PET image reconstruction. The three radioactive point sources were aligned in a horizontal diagonal line within the acrylic glass table. The measurement was repeated with the transducer in four different positions: (A) next to table (edge), (B) next to table (corner), (C) under table, and (D) on table.

Mentions: In order to investigate the artifact effects of the US probe being within the lines of response of the PET detector ring, PET images of three radioactive point sources were acquired while the US probe was positioned close by the sources within the PET field of view. As shown in Figure 4, the three point sources were positioned in a horizontal diagonal line within a fixed small acrylic glass table construction, which was aligned with the laser cross hairs of the PET scanner. Five measurements were performed: one reference measurement without the US transducer and four measurements with the US transducer being fixated at different positions relative to the point sources. The four US probe positions were (a) next to the acrylic glass table in a central position at its edge, (b) next to the acrylic glass table close to a corner, (c) under the acrylic glass table, thus, approximately 6 cm under the central point source, and (d) lying on the table directly over the central point source. For each setup, an attenuation correction CT was acquired so that the US probe was taken into account during PET image reconstruction. The measured activities as well as the relative positions of the three point sources in the reconstructed image were compared, respectively.


First Steps Toward Ultrasound-Based Motion Compensation for Imaging and Therapy: Calibration with an Optical System and 4D PET Imaging.

Schwaab J, Kurz C, Sarti C, Bongers A, Schoenahl F, Bert C, Debus J, Parodi K, Jenne JW - Front Oncol (2015)

Experimental setup for investigation of the US transducer’s influence on PET image reconstruction. The three radioactive point sources were aligned in a horizontal diagonal line within the acrylic glass table. The measurement was repeated with the transducer in four different positions: (A) next to table (edge), (B) next to table (corner), (C) under table, and (D) on table.
© Copyright Policy
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC4663279&req=5

Figure 4: Experimental setup for investigation of the US transducer’s influence on PET image reconstruction. The three radioactive point sources were aligned in a horizontal diagonal line within the acrylic glass table. The measurement was repeated with the transducer in four different positions: (A) next to table (edge), (B) next to table (corner), (C) under table, and (D) on table.
Mentions: In order to investigate the artifact effects of the US probe being within the lines of response of the PET detector ring, PET images of three radioactive point sources were acquired while the US probe was positioned close by the sources within the PET field of view. As shown in Figure 4, the three point sources were positioned in a horizontal diagonal line within a fixed small acrylic glass table construction, which was aligned with the laser cross hairs of the PET scanner. Five measurements were performed: one reference measurement without the US transducer and four measurements with the US transducer being fixated at different positions relative to the point sources. The four US probe positions were (a) next to the acrylic glass table in a central position at its edge, (b) next to the acrylic glass table close to a corner, (c) under the acrylic glass table, thus, approximately 6 cm under the central point source, and (d) lying on the table directly over the central point source. For each setup, an attenuation correction CT was acquired so that the US probe was taken into account during PET image reconstruction. The measured activities as well as the relative positions of the three point sources in the reconstructed image were compared, respectively.

Bottom Line: Furthermore, it is demonstrated that the US probe being within the PET field of view generally has no relevant influence on the image quality.The accuracy and precision of all the steps in the calibration workflow for US tracking-based 4D PET imaging are found to be in an acceptable range for clinical implementation.Eventually, we show in vitro that an US-based motion tracking in absolute room coordinates with a moving US transducer is feasible.

View Article: PubMed Central - PubMed

Affiliation: Mediri GmbH , Heidelberg , Germany.

ABSTRACT
Target motion, particularly in the abdomen, due to respiration or patient movement is still a challenge in many diagnostic and therapeutic processes. Hence, methods to detect and compensate this motion are required. Diagnostic ultrasound (US) represents a non-invasive and dose-free alternative to fluoroscopy, providing more information about internal target motion than respiration belt or optical tracking. The goal of this project is to develop an US-based motion tracking for real-time motion correction in radiation therapy and diagnostic imaging, notably in 4D positron emission tomography (PET). In this work, a workflow is established to enable the transformation of US tracking data to the coordinates of the treatment delivery or imaging system - even if the US probe is moving due to respiration. It is shown that the US tracking signal is equally adequate for 4D PET image reconstruction as the clinically used respiration belt and provides additional opportunities in this concern. Furthermore, it is demonstrated that the US probe being within the PET field of view generally has no relevant influence on the image quality. The accuracy and precision of all the steps in the calibration workflow for US tracking-based 4D PET imaging are found to be in an acceptable range for clinical implementation. Eventually, we show in vitro that an US-based motion tracking in absolute room coordinates with a moving US transducer is feasible.

No MeSH data available.


Related in: MedlinePlus