Characterization of a new commercial single crystal diamond detector for photon- and proton-beam dosimetry.

Akino Y, Gautam A, Coutinho L, Würfel J, Das IJ - J. Radiat. Res. (2015)

Bottom Line: To evaluate the effects of the pre-irradiation, the diamond detector, which had not been irradiated on the day, was set up in the water tank, and the response to 100 MU was measured every 20 s.The profile of the diamond detector was very similar to those of the EDGE and SFD detectors, although the Semiflex and PinPoint chambers showed volume-averaging effects in the penumbrae region.However, at least 1000 cGy of pre-irradiation will be needed for accurate measurements.

View Article: PubMed Central - PubMed

Affiliation: Department of Radiation Oncology, Indiana University School of Medicine, Indianapolis 46202, USA Present address: Division of Health Sciences, Osaka University Graduate School of Medicine, 1-7 Yamadaoka, Suita, Osaka 565-0871, Japan.

No MeSH data available.


Percentage depth–dose of proton beams with the energy range of 8, 16 and 24 cm.
© Copyright Policy - creative-commons
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC4628217&req=5

RRV044F5: Percentage depth–dose of proton beams with the energy range of 8, 16 and 24 cm.

Mentions: Figure 5 illustrates the PDD of the pristine Bragg peak of the proton beam with the energy range of 8, 16 and 24 cm. The data measured with the SCDD showed very similar data, representing the small energy-dependency of the SCDD against the proton beams.Fig. 5.


Characterization of a new commercial single crystal diamond detector for photon- and proton-beam dosimetry.

Akino Y, Gautam A, Coutinho L, Würfel J, Das IJ - J. Radiat. Res. (2015)

Percentage depth–dose of proton beams with the energy range of 8, 16 and 24 cm.
© Copyright Policy - creative-commons
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC4628217&req=5

RRV044F5: Percentage depth–dose of proton beams with the energy range of 8, 16 and 24 cm.
Mentions: Figure 5 illustrates the PDD of the pristine Bragg peak of the proton beam with the energy range of 8, 16 and 24 cm. The data measured with the SCDD showed very similar data, representing the small energy-dependency of the SCDD against the proton beams.Fig. 5.

Bottom Line: To evaluate the effects of the pre-irradiation, the diamond detector, which had not been irradiated on the day, was set up in the water tank, and the response to 100 MU was measured every 20 s.The profile of the diamond detector was very similar to those of the EDGE and SFD detectors, although the Semiflex and PinPoint chambers showed volume-averaging effects in the penumbrae region.However, at least 1000 cGy of pre-irradiation will be needed for accurate measurements.

View Article: PubMed Central - PubMed

Affiliation: Department of Radiation Oncology, Indiana University School of Medicine, Indianapolis 46202, USA Present address: Division of Health Sciences, Osaka University Graduate School of Medicine, 1-7 Yamadaoka, Suita, Osaka 565-0871, Japan.

No MeSH data available.