Water Selective Imaging and bSSFP Banding Artifact Correction in Humans and Small Animals at 3T and 7T, Respectively.

Ribot EJ, Wecker D, Trotier AJ, Dallaudière B, Lefrançois W, Thiaudière E, Franconi JM, Miraux S - PLoS ONE (2015)

Bottom Line: In addition, ankles, full head coverage and legs of volunteers were imaged, demonstrating the possible application of the method also for large FOV.In conclusion, this robust method can be applied in small animals and humans at high magnetic fields.The high SNR and tissue contrast obtained in short acquisition times allows to prescribe bSSFP sequence for several preclinical and clinical applications.

View Article: PubMed Central - PubMed

Affiliation: Centre de Résonance Magnétique des Systèmes Biologiques, UMR 5536, CNRS/University Bordeaux, Bordeaux, France.

ABSTRACT

Introduction: The purpose of this paper is to develop an easy method to generate both fat signal and banding artifact free 3D balanced Steady State Free Precession (bSSFP) images at high magnetic field.

Methods: In order to suppress fat signal and bSSFP banding artifacts, two or four images were acquired with the excitation frequency of the water-selective binomial radiofrequency pulse set On Resonance or shifted by a maximum of 3/4TR. Mice and human volunteers were imaged at 7 T and 3 T, respectively to perform whole-body and musculoskeletal imaging. "Sum-Of-Square" reconstruction was performed and combined or not with parallel imaging.

Results: The frequency selectivity of 1-2-3-2-1 or 1-3-3-1 binomial pulses was preserved after (3/4TR) frequency shifting. Consequently, whole body small animal 3D imaging was performed at 7 T and enabled visualization of small structures within adipose tissue like lymph nodes. In parallel, this method allowed 3D musculoskeletal imaging in humans with high spatial resolution at 3 T. The combination with parallel imaging allowed the acquisition of knee images with ~500 μm resolution images in less than 2 min. In addition, ankles, full head coverage and legs of volunteers were imaged, demonstrating the possible application of the method also for large FOV.

Conclusion: In conclusion, this robust method can be applied in small animals and humans at high magnetic fields. The high SNR and tissue contrast obtained in short acquisition times allows to prescribe bSSFP sequence for several preclinical and clinical applications.

No MeSH data available.


3D coronal images of mouse abdomen acquired at 7T using standard bSSFP or WS-bSSFP sequences.The images shown were either acquired On Resonance (OR), OR + (3/4TR) apart from the resonance (for TR = 4.5ms, the shift equals 166.7Hz), or summed using the SOS technique (SOS4). Arrows point at banding artifacts. Arrowheads indicate the inguinal lymph node location.
© Copyright Policy
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC4591352&req=5

pone.0139249.g002: 3D coronal images of mouse abdomen acquired at 7T using standard bSSFP or WS-bSSFP sequences.The images shown were either acquired On Resonance (OR), OR + (3/4TR) apart from the resonance (for TR = 4.5ms, the shift equals 166.7Hz), or summed using the SOS technique (SOS4). Arrows point at banding artifacts. Arrowheads indicate the inguinal lymph node location.

Mentions: Mouse abdomen imaging was performed at 7T (Fig 2) using a standard bSSFP sequence with a conventional non-binomial RF pulse. Adipose tissues were detected as hyperintense areas (Table 3). Chemical shift artifacts were present particularly between kidneys and visceral fat and around the inguinal lymph nodes (arrowheads). Then, the 3D WS-bSSFP sequence was used. Images acquired with the smaller (0Hz) and the larger (166.7Hz) excitation frequency offsets are shown in Fig 2. In both images, the fat signal was nearly cancelled on the entire FOV (p = 0.004). The banding artifacts were broad (representing between 10 to 24 pixels width), differently located on the two images. Figure A in S2 Fig shows the multiple locations of drop of signals in the mouse kidney in function of the frequency shift. After the SOS 4 reconstruction, the kidney signal was more homogeneous. Quantitatively, the SNR of muscles affected by these signal losses was 4.8±0.5 on the OR images. After applying SOS4 reconstruction, no banding artifact could be detected on the entire 3D WS-bSSFP image (the SNR of muscle affected by banding artifacts was 31.1±6.2, returning to the mean muscle SNR of 29.4±4.7 (p = 0.017)).


Water Selective Imaging and bSSFP Banding Artifact Correction in Humans and Small Animals at 3T and 7T, Respectively.

Ribot EJ, Wecker D, Trotier AJ, Dallaudière B, Lefrançois W, Thiaudière E, Franconi JM, Miraux S - PLoS ONE (2015)

3D coronal images of mouse abdomen acquired at 7T using standard bSSFP or WS-bSSFP sequences.The images shown were either acquired On Resonance (OR), OR + (3/4TR) apart from the resonance (for TR = 4.5ms, the shift equals 166.7Hz), or summed using the SOS technique (SOS4). Arrows point at banding artifacts. Arrowheads indicate the inguinal lymph node location.
© Copyright Policy
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC4591352&req=5

pone.0139249.g002: 3D coronal images of mouse abdomen acquired at 7T using standard bSSFP or WS-bSSFP sequences.The images shown were either acquired On Resonance (OR), OR + (3/4TR) apart from the resonance (for TR = 4.5ms, the shift equals 166.7Hz), or summed using the SOS technique (SOS4). Arrows point at banding artifacts. Arrowheads indicate the inguinal lymph node location.
Mentions: Mouse abdomen imaging was performed at 7T (Fig 2) using a standard bSSFP sequence with a conventional non-binomial RF pulse. Adipose tissues were detected as hyperintense areas (Table 3). Chemical shift artifacts were present particularly between kidneys and visceral fat and around the inguinal lymph nodes (arrowheads). Then, the 3D WS-bSSFP sequence was used. Images acquired with the smaller (0Hz) and the larger (166.7Hz) excitation frequency offsets are shown in Fig 2. In both images, the fat signal was nearly cancelled on the entire FOV (p = 0.004). The banding artifacts were broad (representing between 10 to 24 pixels width), differently located on the two images. Figure A in S2 Fig shows the multiple locations of drop of signals in the mouse kidney in function of the frequency shift. After the SOS 4 reconstruction, the kidney signal was more homogeneous. Quantitatively, the SNR of muscles affected by these signal losses was 4.8±0.5 on the OR images. After applying SOS4 reconstruction, no banding artifact could be detected on the entire 3D WS-bSSFP image (the SNR of muscle affected by banding artifacts was 31.1±6.2, returning to the mean muscle SNR of 29.4±4.7 (p = 0.017)).

Bottom Line: In addition, ankles, full head coverage and legs of volunteers were imaged, demonstrating the possible application of the method also for large FOV.In conclusion, this robust method can be applied in small animals and humans at high magnetic fields.The high SNR and tissue contrast obtained in short acquisition times allows to prescribe bSSFP sequence for several preclinical and clinical applications.

View Article: PubMed Central - PubMed

Affiliation: Centre de Résonance Magnétique des Systèmes Biologiques, UMR 5536, CNRS/University Bordeaux, Bordeaux, France.

ABSTRACT

Introduction: The purpose of this paper is to develop an easy method to generate both fat signal and banding artifact free 3D balanced Steady State Free Precession (bSSFP) images at high magnetic field.

Methods: In order to suppress fat signal and bSSFP banding artifacts, two or four images were acquired with the excitation frequency of the water-selective binomial radiofrequency pulse set On Resonance or shifted by a maximum of 3/4TR. Mice and human volunteers were imaged at 7 T and 3 T, respectively to perform whole-body and musculoskeletal imaging. "Sum-Of-Square" reconstruction was performed and combined or not with parallel imaging.

Results: The frequency selectivity of 1-2-3-2-1 or 1-3-3-1 binomial pulses was preserved after (3/4TR) frequency shifting. Consequently, whole body small animal 3D imaging was performed at 7 T and enabled visualization of small structures within adipose tissue like lymph nodes. In parallel, this method allowed 3D musculoskeletal imaging in humans with high spatial resolution at 3 T. The combination with parallel imaging allowed the acquisition of knee images with ~500 μm resolution images in less than 2 min. In addition, ankles, full head coverage and legs of volunteers were imaged, demonstrating the possible application of the method also for large FOV.

Conclusion: In conclusion, this robust method can be applied in small animals and humans at high magnetic fields. The high SNR and tissue contrast obtained in short acquisition times allows to prescribe bSSFP sequence for several preclinical and clinical applications.

No MeSH data available.