A review of 3D first-pass, whole-heart, myocardial perfusion cardiovascular magnetic resonance.

Fair MJ, Gatehouse PD, DiBella EV, Firmin DN - J Cardiovasc Magn Reson (2015)

Bottom Line: The mechanisms include rapid sequences, non-Cartesian k-space trajectories, reduced k-space acquisitions, parallel imaging reconstructions and compressed sensing.An attempt is made to explain, rather than simply state, the varying methods with the hope that it will give an appreciation of the different components making up a 3D FPP protocol.Basic estimates demonstrating the required total acceleration factors in typical 3D FPP cases are included, providing context for the extent that each acceleration method can contribute to the required imaging speed, as well as potential limitations in present 3D FPP literature.

View Article: PubMed Central - PubMed

Affiliation: National Heart & Lung Institute, Imperial College London, London, UK. M.Fair@rbht.nhs.uk.

ABSTRACT
A comprehensive review is undertaken of the methods available for 3D whole-heart first-pass perfusion (FPP) and their application to date, with particular focus on possible acceleration techniques. Following a summary of the parameters typically desired of 3D FPP methods, the review explains the mechanisms of key acceleration techniques and their potential use in FPP for attaining 3D acquisitions. The mechanisms include rapid sequences, non-Cartesian k-space trajectories, reduced k-space acquisitions, parallel imaging reconstructions and compressed sensing. An attempt is made to explain, rather than simply state, the varying methods with the hope that it will give an appreciation of the different components making up a 3D FPP protocol. Basic estimates demonstrating the required total acceleration factors in typical 3D FPP cases are included, providing context for the extent that each acceleration method can contribute to the required imaging speed, as well as potential limitations in present 3D FPP literature. Although many 3D FPP methods are too early in development for the type of clinical trials required to show any clear benefit over current 2D FPP methods, the review includes the small but growing quantity of clinical research work already using 3D FPP, alongside the more technical work. Broader challenges concerning FPP such as quantitative analysis are not covered, but challenges with particular impact on 3D FPP methods, particularly with regards to motion effects, are discussed along with anticipated future work in the field.

No MeSH data available.


Related in: MedlinePlus

Non-Cartesian trajectories. Examples of three potential alternate trajectories discussed in the text. EPI (a) demonstrated with an ETL of 4, a spiral trajectory (b) with 4 interleaves and a radial projection design (c). Partition encoding direction in (b) and (c) is the same as for (a)
© Copyright Policy - open-access
Related In: Results  -  Collection

License 1 - License 2
getmorefigures.php?uid=PMC4522116&req=5

Fig2: Non-Cartesian trajectories. Examples of three potential alternate trajectories discussed in the text. EPI (a) demonstrated with an ETL of 4, a spiral trajectory (b) with 4 interleaves and a radial projection design (c). Partition encoding direction in (b) and (c) is the same as for (a)

Mentions: Echo-planar Imaging (EPI) is geometrically closest to typical Cartesian k-space coverage (Fig. 2a) [23]. The premise is to collect multiple ‘lines’ of raw data by a series of gradient echoes after each RF pulse, allowing acceleration by omitting RF excitations. Due mainly to the limited echo-train length (ETL) achievable as a consequence of cardiac motion and main-field inhomogeneities around the heart, hybrid EPI (h-EPI) is generally used in FPP [26] with limited applicability of true single-shot EPI [27, 28]. The compromise for h-EPI is made between increasing the ETL – which reduces total image time – and the corresponding increased unreliability [26, 29] which also increases with main field strength. Despite this, h-EPI in 2D FPP typically uses an ETL of around 4 at 1.5 T (no current examples at 3 T), corresponding to acceleration factors of approximately 2 compared to the FLASH timings calculated earlier (see Appendix). Early examples of 2D FPP with extended LV coverage used h-EPI [26, 30, 31] but so far 3D EPI imaging has largely been limited to non-cardiac work.Fig. 2


A review of 3D first-pass, whole-heart, myocardial perfusion cardiovascular magnetic resonance.

Fair MJ, Gatehouse PD, DiBella EV, Firmin DN - J Cardiovasc Magn Reson (2015)

Non-Cartesian trajectories. Examples of three potential alternate trajectories discussed in the text. EPI (a) demonstrated with an ETL of 4, a spiral trajectory (b) with 4 interleaves and a radial projection design (c). Partition encoding direction in (b) and (c) is the same as for (a)
© Copyright Policy - open-access
Related In: Results  -  Collection

License 1 - License 2
Show All Figures
getmorefigures.php?uid=PMC4522116&req=5

Fig2: Non-Cartesian trajectories. Examples of three potential alternate trajectories discussed in the text. EPI (a) demonstrated with an ETL of 4, a spiral trajectory (b) with 4 interleaves and a radial projection design (c). Partition encoding direction in (b) and (c) is the same as for (a)
Mentions: Echo-planar Imaging (EPI) is geometrically closest to typical Cartesian k-space coverage (Fig. 2a) [23]. The premise is to collect multiple ‘lines’ of raw data by a series of gradient echoes after each RF pulse, allowing acceleration by omitting RF excitations. Due mainly to the limited echo-train length (ETL) achievable as a consequence of cardiac motion and main-field inhomogeneities around the heart, hybrid EPI (h-EPI) is generally used in FPP [26] with limited applicability of true single-shot EPI [27, 28]. The compromise for h-EPI is made between increasing the ETL – which reduces total image time – and the corresponding increased unreliability [26, 29] which also increases with main field strength. Despite this, h-EPI in 2D FPP typically uses an ETL of around 4 at 1.5 T (no current examples at 3 T), corresponding to acceleration factors of approximately 2 compared to the FLASH timings calculated earlier (see Appendix). Early examples of 2D FPP with extended LV coverage used h-EPI [26, 30, 31] but so far 3D EPI imaging has largely been limited to non-cardiac work.Fig. 2

Bottom Line: The mechanisms include rapid sequences, non-Cartesian k-space trajectories, reduced k-space acquisitions, parallel imaging reconstructions and compressed sensing.An attempt is made to explain, rather than simply state, the varying methods with the hope that it will give an appreciation of the different components making up a 3D FPP protocol.Basic estimates demonstrating the required total acceleration factors in typical 3D FPP cases are included, providing context for the extent that each acceleration method can contribute to the required imaging speed, as well as potential limitations in present 3D FPP literature.

View Article: PubMed Central - PubMed

Affiliation: National Heart & Lung Institute, Imperial College London, London, UK. M.Fair@rbht.nhs.uk.

ABSTRACT
A comprehensive review is undertaken of the methods available for 3D whole-heart first-pass perfusion (FPP) and their application to date, with particular focus on possible acceleration techniques. Following a summary of the parameters typically desired of 3D FPP methods, the review explains the mechanisms of key acceleration techniques and their potential use in FPP for attaining 3D acquisitions. The mechanisms include rapid sequences, non-Cartesian k-space trajectories, reduced k-space acquisitions, parallel imaging reconstructions and compressed sensing. An attempt is made to explain, rather than simply state, the varying methods with the hope that it will give an appreciation of the different components making up a 3D FPP protocol. Basic estimates demonstrating the required total acceleration factors in typical 3D FPP cases are included, providing context for the extent that each acceleration method can contribute to the required imaging speed, as well as potential limitations in present 3D FPP literature. Although many 3D FPP methods are too early in development for the type of clinical trials required to show any clear benefit over current 2D FPP methods, the review includes the small but growing quantity of clinical research work already using 3D FPP, alongside the more technical work. Broader challenges concerning FPP such as quantitative analysis are not covered, but challenges with particular impact on 3D FPP methods, particularly with regards to motion effects, are discussed along with anticipated future work in the field.

No MeSH data available.


Related in: MedlinePlus