The Impact of Blood Rheology on Drug Transport in Stented Arteries: Steady Simulations.

Vijayaratnam PR, O'Brien CC, Reizes JA, Barber TJ, Edelman ER - PLoS ONE (2015)

Bottom Line: Recirculation lengths were found to reduce by as much as 24% with the inclusion of a non-Newtonian rheological model.Despite the small sensitivity to the effects of viscosity variations, the spatial distribution of drug matter in the tissue was found to be significantly affected by rheological model selection.The clinical significance of these results is that they convey that the magnitude of drug uptake in stent-based drug delivery is relatively insensitive to individual variations in blood rheology.

View Article: PubMed Central - PubMed

Affiliation: School of Mechanical and Manufacturing Engineering, the University of New South Wales, Sydney, New South Wales, Australia.

ABSTRACT

Background and methods: It is important to ensure that blood flow is modelled accurately in numerical studies of arteries featuring drug-eluting stents due to the significant proportion of drug transport from the stent into the arterial wall which is flow-mediated. Modelling blood is complicated, however, by variations in blood rheological behaviour between individuals, blood's complex near-wall behaviour, and the large number of rheological models which have been proposed. In this study, a series of steady-state computational fluid dynamics analyses were performed in which the traditional Newtonian model was compared against a range of non-Newtonian models. The impact of these rheological models was elucidated through comparisons of haemodynamic flow details and drug transport behaviour at various blood flow rates.

Results: Recirculation lengths were found to reduce by as much as 24% with the inclusion of a non-Newtonian rheological model. Another model possessing the viscosity and density of blood plasma was also implemented to account for near-wall red blood cell losses and yielded recirculation length increases of up to 59%. However, the deviation from the average drug concentration in the tissue obtained with the Newtonian model was observed to be less than 5% in all cases except one. Despite the small sensitivity to the effects of viscosity variations, the spatial distribution of drug matter in the tissue was found to be significantly affected by rheological model selection.

Conclusions/significance: These results may be used to guide blood rheological model selection in future numerical studies. The clinical significance of these results is that they convey that the magnitude of drug uptake in stent-based drug delivery is relatively insensitive to individual variations in blood rheology. Furthermore, the finding that flow separation regions formed downstream of the stent struts diminish drug uptake may be of interest to device designers.

No MeSH data available.


Related in: MedlinePlus

Experimental measurements of blood viscosity and non-Newtonian blood rheological models.
© Copyright Policy
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC4466567&req=5

pone.0128178.g002: Experimental measurements of blood viscosity and non-Newtonian blood rheological models.

Mentions: In the Newtonian fluid model, the dynamic viscosity is assumed to remain at a constant value of = 0.00345 Pa·s [7]. Although this assumption greatly simplifies the modelling of blood, it has been found to only be acceptable in flows in which strain rates above 100s-1 are encountered [24]. Such high strain rates are found in larger arteries and consequently many investigators have justified their assumption of blood being a Newtonian fluid by emphasising the large size of the arteries that they were modelling [1, 3]. However, the separated flow regions near the stent are characterised by low velocities, meaning that strain rates are modest. As non-Newtonian behaviour may significantly affect velocity distributions in these regions, it may also affect the rate at which drug is removed from the surface and convected by blood. Hence, five non-Newtonian models of blood rheology which incorporated shear- and haematocrit-dependent viscosities were also examined in this study. These models were chosen because they span the spectrum of viscosity variations with respect to strain-rate which have been observed in clinical data (Fig 2 [25–34]). The mathematical formulations of these models are given in Table 1.


The Impact of Blood Rheology on Drug Transport in Stented Arteries: Steady Simulations.

Vijayaratnam PR, O'Brien CC, Reizes JA, Barber TJ, Edelman ER - PLoS ONE (2015)

Experimental measurements of blood viscosity and non-Newtonian blood rheological models.
© Copyright Policy
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC4466567&req=5

pone.0128178.g002: Experimental measurements of blood viscosity and non-Newtonian blood rheological models.
Mentions: In the Newtonian fluid model, the dynamic viscosity is assumed to remain at a constant value of = 0.00345 Pa·s [7]. Although this assumption greatly simplifies the modelling of blood, it has been found to only be acceptable in flows in which strain rates above 100s-1 are encountered [24]. Such high strain rates are found in larger arteries and consequently many investigators have justified their assumption of blood being a Newtonian fluid by emphasising the large size of the arteries that they were modelling [1, 3]. However, the separated flow regions near the stent are characterised by low velocities, meaning that strain rates are modest. As non-Newtonian behaviour may significantly affect velocity distributions in these regions, it may also affect the rate at which drug is removed from the surface and convected by blood. Hence, five non-Newtonian models of blood rheology which incorporated shear- and haematocrit-dependent viscosities were also examined in this study. These models were chosen because they span the spectrum of viscosity variations with respect to strain-rate which have been observed in clinical data (Fig 2 [25–34]). The mathematical formulations of these models are given in Table 1.

Bottom Line: Recirculation lengths were found to reduce by as much as 24% with the inclusion of a non-Newtonian rheological model.Despite the small sensitivity to the effects of viscosity variations, the spatial distribution of drug matter in the tissue was found to be significantly affected by rheological model selection.The clinical significance of these results is that they convey that the magnitude of drug uptake in stent-based drug delivery is relatively insensitive to individual variations in blood rheology.

View Article: PubMed Central - PubMed

Affiliation: School of Mechanical and Manufacturing Engineering, the University of New South Wales, Sydney, New South Wales, Australia.

ABSTRACT

Background and methods: It is important to ensure that blood flow is modelled accurately in numerical studies of arteries featuring drug-eluting stents due to the significant proportion of drug transport from the stent into the arterial wall which is flow-mediated. Modelling blood is complicated, however, by variations in blood rheological behaviour between individuals, blood's complex near-wall behaviour, and the large number of rheological models which have been proposed. In this study, a series of steady-state computational fluid dynamics analyses were performed in which the traditional Newtonian model was compared against a range of non-Newtonian models. The impact of these rheological models was elucidated through comparisons of haemodynamic flow details and drug transport behaviour at various blood flow rates.

Results: Recirculation lengths were found to reduce by as much as 24% with the inclusion of a non-Newtonian rheological model. Another model possessing the viscosity and density of blood plasma was also implemented to account for near-wall red blood cell losses and yielded recirculation length increases of up to 59%. However, the deviation from the average drug concentration in the tissue obtained with the Newtonian model was observed to be less than 5% in all cases except one. Despite the small sensitivity to the effects of viscosity variations, the spatial distribution of drug matter in the tissue was found to be significantly affected by rheological model selection.

Conclusions/significance: These results may be used to guide blood rheological model selection in future numerical studies. The clinical significance of these results is that they convey that the magnitude of drug uptake in stent-based drug delivery is relatively insensitive to individual variations in blood rheology. Furthermore, the finding that flow separation regions formed downstream of the stent struts diminish drug uptake may be of interest to device designers.