Evaluating the accuracy of diffusion MRI models in white matter.

Rokem A, Yeatman JD, Pestilli F, Kay KN, Mezer A, van der Walt S, Wandell BA - PLoS ONE (2015)

Bottom Line: This is the first evaluation of model-accuracy of these models.In most of the white matter the DTM predicts the data more accurately than test-retest reliability; SFM model-accuracy is higher than test-retest reliability and also higher than the DTM model-accuracy, particularly for measurements with (a) a b-value above 1000 in locations containing fiber crossings, and (b) in the regions of the brain surrounding the optic radiations.The SFM also has better parameter-validity: it more accurately estimates the fiber orientation distribution function (fODF) in each voxel, which is useful for fiber tracking.

View Article: PubMed Central - PubMed

Affiliation: Department of Psychology, Stanford, Stanford, California, United States of America.

ABSTRACT
Models of diffusion MRI within a voxel are useful for making inferences about the properties of the tissue and inferring fiber orientation distribution used by tractography algorithms. A useful model must fit the data accurately. However, evaluations of model-accuracy of commonly used models have not been published before. Here, we evaluate model-accuracy of the two main classes of diffusion MRI models. The diffusion tensor model (DTM) summarizes diffusion as a 3-dimensional Gaussian distribution. Sparse fascicle models (SFM) summarize the signal as a sum of signals originating from a collection of fascicles oriented in different directions. We use cross-validation to assess model-accuracy at different gradient amplitudes (b-values) throughout the white matter. Specifically, we fit each model to all the white matter voxels in one data set and then use the model to predict a second, independent data set. This is the first evaluation of model-accuracy of these models. In most of the white matter the DTM predicts the data more accurately than test-retest reliability; SFM model-accuracy is higher than test-retest reliability and also higher than the DTM model-accuracy, particularly for measurements with (a) a b-value above 1000 in locations containing fiber crossings, and (b) in the regions of the brain surrounding the optic radiations. The SFM also has better parameter-validity: it more accurately estimates the fiber orientation distribution function (fODF) in each voxel, which is useful for fiber tracking.

Show MeSH
RMSE and SNR of diffusion MRI measurements.Error bars delineate the 95% interquantile range. RMSE does not change across b-values, but SNR changes substantially, with the median decreasing from approximately 7 (b = 1000 s/mm2) to approximately 2 (b = 2000 s/mm2).
© Copyright Policy
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC4400066&req=5

pone.0123272.g002: RMSE and SNR of diffusion MRI measurements.Error bars delineate the 95% interquantile range. RMSE does not change across b-values, but SNR changes substantially, with the median decreasing from approximately 7 (b = 1000 s/mm2) to approximately 2 (b = 2000 s/mm2).

Mentions: To estimate the goodness-of-fit of a model we use cross-validation. Specifically, we estimate the parameters of the model on one set of data; we then use the model parameters to predict the signal in a second, independent data set. We use RMSE of the signal as a measure of model accuracy. We further assess model-accuracy by comparing the model RMSE to the RMSE of repeated measurements: test-retest reliability. DWI data was collected twice in each b-value and test-retest reliability is calculated as the RMSE between the two measurements in each voxel across directions of measurement. RMSE is given in the units of the measurement. In the case of MRI data, these are the scanner signal units, which do not have a straightforward physical interpretation. Thus, it is difficult to compare RMSE values across different locations in the brain and across different measurement parameters. The distribution of RMSE of test-retest reliability is very similar across b-values (see Results, Fig 2). We speculate that this indicates that the noise arises principally from sources that are independent of the diffusion itself, including subject motion, thermal changes in the scanner equipment, and physiological noise. Nevertheless, RMSE does not provide a natural benchmark. Less error is better, but it is not clear how small of an error is good enough.


Evaluating the accuracy of diffusion MRI models in white matter.

Rokem A, Yeatman JD, Pestilli F, Kay KN, Mezer A, van der Walt S, Wandell BA - PLoS ONE (2015)

RMSE and SNR of diffusion MRI measurements.Error bars delineate the 95% interquantile range. RMSE does not change across b-values, but SNR changes substantially, with the median decreasing from approximately 7 (b = 1000 s/mm2) to approximately 2 (b = 2000 s/mm2).
© Copyright Policy
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC4400066&req=5

pone.0123272.g002: RMSE and SNR of diffusion MRI measurements.Error bars delineate the 95% interquantile range. RMSE does not change across b-values, but SNR changes substantially, with the median decreasing from approximately 7 (b = 1000 s/mm2) to approximately 2 (b = 2000 s/mm2).
Mentions: To estimate the goodness-of-fit of a model we use cross-validation. Specifically, we estimate the parameters of the model on one set of data; we then use the model parameters to predict the signal in a second, independent data set. We use RMSE of the signal as a measure of model accuracy. We further assess model-accuracy by comparing the model RMSE to the RMSE of repeated measurements: test-retest reliability. DWI data was collected twice in each b-value and test-retest reliability is calculated as the RMSE between the two measurements in each voxel across directions of measurement. RMSE is given in the units of the measurement. In the case of MRI data, these are the scanner signal units, which do not have a straightforward physical interpretation. Thus, it is difficult to compare RMSE values across different locations in the brain and across different measurement parameters. The distribution of RMSE of test-retest reliability is very similar across b-values (see Results, Fig 2). We speculate that this indicates that the noise arises principally from sources that are independent of the diffusion itself, including subject motion, thermal changes in the scanner equipment, and physiological noise. Nevertheless, RMSE does not provide a natural benchmark. Less error is better, but it is not clear how small of an error is good enough.

Bottom Line: This is the first evaluation of model-accuracy of these models.In most of the white matter the DTM predicts the data more accurately than test-retest reliability; SFM model-accuracy is higher than test-retest reliability and also higher than the DTM model-accuracy, particularly for measurements with (a) a b-value above 1000 in locations containing fiber crossings, and (b) in the regions of the brain surrounding the optic radiations.The SFM also has better parameter-validity: it more accurately estimates the fiber orientation distribution function (fODF) in each voxel, which is useful for fiber tracking.

View Article: PubMed Central - PubMed

Affiliation: Department of Psychology, Stanford, Stanford, California, United States of America.

ABSTRACT
Models of diffusion MRI within a voxel are useful for making inferences about the properties of the tissue and inferring fiber orientation distribution used by tractography algorithms. A useful model must fit the data accurately. However, evaluations of model-accuracy of commonly used models have not been published before. Here, we evaluate model-accuracy of the two main classes of diffusion MRI models. The diffusion tensor model (DTM) summarizes diffusion as a 3-dimensional Gaussian distribution. Sparse fascicle models (SFM) summarize the signal as a sum of signals originating from a collection of fascicles oriented in different directions. We use cross-validation to assess model-accuracy at different gradient amplitudes (b-values) throughout the white matter. Specifically, we fit each model to all the white matter voxels in one data set and then use the model to predict a second, independent data set. This is the first evaluation of model-accuracy of these models. In most of the white matter the DTM predicts the data more accurately than test-retest reliability; SFM model-accuracy is higher than test-retest reliability and also higher than the DTM model-accuracy, particularly for measurements with (a) a b-value above 1000 in locations containing fiber crossings, and (b) in the regions of the brain surrounding the optic radiations. The SFM also has better parameter-validity: it more accurately estimates the fiber orientation distribution function (fODF) in each voxel, which is useful for fiber tracking.