Interaction mechanisms of cavitation bubbles induced by spatially and temporally separated fs-laser pulses.

Tinne N, Kaune B, Krüger A, Ripken T - PLoS ONE (2014)

Bottom Line: The emerging use of femtosecond lasers with high repetition rates in the MHz regime together with limited scan speed implies possible mutual optical and dynamical interaction effects of the individual cutting spots.Furthermore, the overall efficiency of energy conversion into controlled mechanical impact should be maximized compared to the transmitted pulse energy and unwanted long range mechanical side effects, e.g. shock waves, axial jet components.In conclusion, these experimental results are of great importance for the prospective optimization of the ophthalmic surgical process with high-repetition rate fs-lasers.

View Article: PubMed Central - PubMed

Affiliation: Laser Zentrum Hannover e.V., Biomedical Optics Department, Hannover, Germany.

ABSTRACT
The emerging use of femtosecond lasers with high repetition rates in the MHz regime together with limited scan speed implies possible mutual optical and dynamical interaction effects of the individual cutting spots. In order to get more insight into the dynamics a time-resolved photographic analysis of the interaction of cavitation bubbles is presented. Particularly, we investigated the influence of fs-laser pulses and their resulting bubble dynamics with various spatial as well as temporal separations. Different time courses of characteristic interaction effects between the cavitation bubbles were observed depending on pulse energy and spatio-temporal pulse separation. These ranged from merely no interaction to the phenomena of strong water jet formation. Afterwards, the mechanisms are discussed regarding their impact on the medical application of effective tissue cutting lateral to the laser beam direction with best possible axial precision: the mechanical forces of photodisruption as well as the occurring water jet should have low axial extend and a preferably lateral priority. Furthermore, the overall efficiency of energy conversion into controlled mechanical impact should be maximized compared to the transmitted pulse energy and unwanted long range mechanical side effects, e.g. shock waves, axial jet components. In conclusion, these experimental results are of great importance for the prospective optimization of the ophthalmic surgical process with high-repetition rate fs-lasers.

Show MeSH

Related in: MedlinePlus

Schematic depiction of the experimental setup of the laser path (red) and the illumination path (orange).Single pulses of the fs-laser are selected by an acousto-optic modulator (AOM), half-wave plate and polarizing beam-splitter cube allow for laser power adjustment. Subsequent laser pulses are spatially separated via polygon scanner and a Keplerian telescope imaging (see also magnified image detail). The focal region inside the sample medium-filled cuvette is illuminated homogeneously by Koehler illumination and a magnified image of the cavitation bubble is reproduced on the chip of the CCD camera.
© Copyright Policy
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC4263672&req=5

pone-0114437-g001: Schematic depiction of the experimental setup of the laser path (red) and the illumination path (orange).Single pulses of the fs-laser are selected by an acousto-optic modulator (AOM), half-wave plate and polarizing beam-splitter cube allow for laser power adjustment. Subsequent laser pulses are spatially separated via polygon scanner and a Keplerian telescope imaging (see also magnified image detail). The focal region inside the sample medium-filled cuvette is illuminated homogeneously by Koehler illumination and a magnified image of the cavitation bubble is reproduced on the chip of the CCD camera.

Mentions: The experimental setup which was an installation for time-resolved photography of the cavitation bubble dynamics can be divided in two light paths; it is shown schematically in Fig. 1. The red one is the path of the fs-laser beam with finally focusing it into a cuvette in order to create the LIOB. The orange beam is used for illumination as well as imaging of the cavitation bubbles.


Interaction mechanisms of cavitation bubbles induced by spatially and temporally separated fs-laser pulses.

Tinne N, Kaune B, Krüger A, Ripken T - PLoS ONE (2014)

Schematic depiction of the experimental setup of the laser path (red) and the illumination path (orange).Single pulses of the fs-laser are selected by an acousto-optic modulator (AOM), half-wave plate and polarizing beam-splitter cube allow for laser power adjustment. Subsequent laser pulses are spatially separated via polygon scanner and a Keplerian telescope imaging (see also magnified image detail). The focal region inside the sample medium-filled cuvette is illuminated homogeneously by Koehler illumination and a magnified image of the cavitation bubble is reproduced on the chip of the CCD camera.
© Copyright Policy
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC4263672&req=5

pone-0114437-g001: Schematic depiction of the experimental setup of the laser path (red) and the illumination path (orange).Single pulses of the fs-laser are selected by an acousto-optic modulator (AOM), half-wave plate and polarizing beam-splitter cube allow for laser power adjustment. Subsequent laser pulses are spatially separated via polygon scanner and a Keplerian telescope imaging (see also magnified image detail). The focal region inside the sample medium-filled cuvette is illuminated homogeneously by Koehler illumination and a magnified image of the cavitation bubble is reproduced on the chip of the CCD camera.
Mentions: The experimental setup which was an installation for time-resolved photography of the cavitation bubble dynamics can be divided in two light paths; it is shown schematically in Fig. 1. The red one is the path of the fs-laser beam with finally focusing it into a cuvette in order to create the LIOB. The orange beam is used for illumination as well as imaging of the cavitation bubbles.

Bottom Line: The emerging use of femtosecond lasers with high repetition rates in the MHz regime together with limited scan speed implies possible mutual optical and dynamical interaction effects of the individual cutting spots.Furthermore, the overall efficiency of energy conversion into controlled mechanical impact should be maximized compared to the transmitted pulse energy and unwanted long range mechanical side effects, e.g. shock waves, axial jet components.In conclusion, these experimental results are of great importance for the prospective optimization of the ophthalmic surgical process with high-repetition rate fs-lasers.

View Article: PubMed Central - PubMed

Affiliation: Laser Zentrum Hannover e.V., Biomedical Optics Department, Hannover, Germany.

ABSTRACT
The emerging use of femtosecond lasers with high repetition rates in the MHz regime together with limited scan speed implies possible mutual optical and dynamical interaction effects of the individual cutting spots. In order to get more insight into the dynamics a time-resolved photographic analysis of the interaction of cavitation bubbles is presented. Particularly, we investigated the influence of fs-laser pulses and their resulting bubble dynamics with various spatial as well as temporal separations. Different time courses of characteristic interaction effects between the cavitation bubbles were observed depending on pulse energy and spatio-temporal pulse separation. These ranged from merely no interaction to the phenomena of strong water jet formation. Afterwards, the mechanisms are discussed regarding their impact on the medical application of effective tissue cutting lateral to the laser beam direction with best possible axial precision: the mechanical forces of photodisruption as well as the occurring water jet should have low axial extend and a preferably lateral priority. Furthermore, the overall efficiency of energy conversion into controlled mechanical impact should be maximized compared to the transmitted pulse energy and unwanted long range mechanical side effects, e.g. shock waves, axial jet components. In conclusion, these experimental results are of great importance for the prospective optimization of the ophthalmic surgical process with high-repetition rate fs-lasers.

Show MeSH
Related in: MedlinePlus