Pore Geometry Optimization of Titanium (Ti6Al4V) Alloy, for Its Application in the Fabrication of Customized Hip Implants.

Roy S, Panda D, Khutia N, Chowdhury AR - Int J Biomater (2014)

Bottom Line: The present study investigates the mechanical response of representative volume elements of porous Ti-6Al-4V alloy, to arrive at a desired range of pore geometries that would optimize the reduction in stiffness necessary for biocompatibility with the stress concentration arising around the pore periphery, under physiological loading conditions with respect to orthopedic hip implants.To perform a detailed analysis of the response of the porous structure over the entire spectrum of loading conditions that a hip implant is subjected to in vivo, the mechanical responses of 3D finite element models of cubic and rectangular parallelepiped geometries, with porosities varying over a range of 10% to 60%, are simulated under representative compressive, flexural as well as combined loading conditions.The results that are obtained are used to suggest a range of pore diameters that lower the effective stiffness and modulus of the implant to around 60% of the stiffness and modulus of dense solid implants while keeping the stress levels within permissible limits.

View Article: PubMed Central - PubMed

Affiliation: Department of Aerospace Engineering and Applied Mechanics, Indian Institute of Engineering Science and Technology, Shibpur 711103, India.

ABSTRACT
The present study investigates the mechanical response of representative volume elements of porous Ti-6Al-4V alloy, to arrive at a desired range of pore geometries that would optimize the reduction in stiffness necessary for biocompatibility with the stress concentration arising around the pore periphery, under physiological loading conditions with respect to orthopedic hip implants. A comparative study of the two is performed with the aid of a newly defined optimizing parameter called pore efficiency that takes into consideration both the stiffness quantity and the stress localization around pores. To perform a detailed analysis of the response of the porous structure over the entire spectrum of loading conditions that a hip implant is subjected to in vivo, the mechanical responses of 3D finite element models of cubic and rectangular parallelepiped geometries, with porosities varying over a range of 10% to 60%, are simulated under representative compressive, flexural as well as combined loading conditions. The results that are obtained are used to suggest a range of pore diameters that lower the effective stiffness and modulus of the implant to around 60% of the stiffness and modulus of dense solid implants while keeping the stress levels within permissible limits.

No MeSH data available.


Related in: MedlinePlus

Plots depicting variation of pore efficiency under combined loading, μo, against pore diameter for different percentages of porosity of 3D porous FE cantilever beam models subjected to combined loading action.
© Copyright Policy - open-access
Related In: Results  -  Collection


getmorefigures.php?uid=PMC4221907&req=5

fig16: Plots depicting variation of pore efficiency under combined loading, μo, against pore diameter for different percentages of porosity of 3D porous FE cantilever beam models subjected to combined loading action.

Mentions: Finally, in order to optimize the effects of decreasing overall stiffness with increasing stress concentration against porosity, a new optimizing parameter similar to the one introduced for flexural loading response, called pore efficiency under combined loading, μo = 1/(Koσo) is introduced and examined in Figure 16. As the expression suggests, a higher value of pore efficiency indicates a greater suitability of a particular pore diameter over the other diameters. The plots of this parameter for different porosities (Figure 16) reveal that, for a given porosity, the pore efficiency varies appreciably with pore diameter, attaining their minimum values for pore diameters in the proximity of 3 mm for the higher porosities of 30% to 60%. But as the pore diameter is varied across different porosities, the values of maximum pore efficiency for a given porosity decrease steadily from the porosity of 10% to 60%. But since the decrement of the same with increasing porosity is not appreciably steep, it may be inferred that the conflicting effects of decreasing stiffness and increasing stress concentration compensate each other to a certain extent to keep the value of their product, as used in the optimizing parameter, from increasing by a substantial amount. As our primary objective lies in minimizing the stiffness of the implant material without letting the stress concentration exceed allowable limits, a check needs to be performed so as to determine whether maximum stiffness reduction for the higher porosities of 35% and 60% can be achieved while keeping the maximum von Mises stress values for combined loading within permissible limits.


Pore Geometry Optimization of Titanium (Ti6Al4V) Alloy, for Its Application in the Fabrication of Customized Hip Implants.

Roy S, Panda D, Khutia N, Chowdhury AR - Int J Biomater (2014)

Plots depicting variation of pore efficiency under combined loading, μo, against pore diameter for different percentages of porosity of 3D porous FE cantilever beam models subjected to combined loading action.
© Copyright Policy - open-access
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC4221907&req=5

fig16: Plots depicting variation of pore efficiency under combined loading, μo, against pore diameter for different percentages of porosity of 3D porous FE cantilever beam models subjected to combined loading action.
Mentions: Finally, in order to optimize the effects of decreasing overall stiffness with increasing stress concentration against porosity, a new optimizing parameter similar to the one introduced for flexural loading response, called pore efficiency under combined loading, μo = 1/(Koσo) is introduced and examined in Figure 16. As the expression suggests, a higher value of pore efficiency indicates a greater suitability of a particular pore diameter over the other diameters. The plots of this parameter for different porosities (Figure 16) reveal that, for a given porosity, the pore efficiency varies appreciably with pore diameter, attaining their minimum values for pore diameters in the proximity of 3 mm for the higher porosities of 30% to 60%. But as the pore diameter is varied across different porosities, the values of maximum pore efficiency for a given porosity decrease steadily from the porosity of 10% to 60%. But since the decrement of the same with increasing porosity is not appreciably steep, it may be inferred that the conflicting effects of decreasing stiffness and increasing stress concentration compensate each other to a certain extent to keep the value of their product, as used in the optimizing parameter, from increasing by a substantial amount. As our primary objective lies in minimizing the stiffness of the implant material without letting the stress concentration exceed allowable limits, a check needs to be performed so as to determine whether maximum stiffness reduction for the higher porosities of 35% and 60% can be achieved while keeping the maximum von Mises stress values for combined loading within permissible limits.

Bottom Line: The present study investigates the mechanical response of representative volume elements of porous Ti-6Al-4V alloy, to arrive at a desired range of pore geometries that would optimize the reduction in stiffness necessary for biocompatibility with the stress concentration arising around the pore periphery, under physiological loading conditions with respect to orthopedic hip implants.To perform a detailed analysis of the response of the porous structure over the entire spectrum of loading conditions that a hip implant is subjected to in vivo, the mechanical responses of 3D finite element models of cubic and rectangular parallelepiped geometries, with porosities varying over a range of 10% to 60%, are simulated under representative compressive, flexural as well as combined loading conditions.The results that are obtained are used to suggest a range of pore diameters that lower the effective stiffness and modulus of the implant to around 60% of the stiffness and modulus of dense solid implants while keeping the stress levels within permissible limits.

View Article: PubMed Central - PubMed

Affiliation: Department of Aerospace Engineering and Applied Mechanics, Indian Institute of Engineering Science and Technology, Shibpur 711103, India.

ABSTRACT
The present study investigates the mechanical response of representative volume elements of porous Ti-6Al-4V alloy, to arrive at a desired range of pore geometries that would optimize the reduction in stiffness necessary for biocompatibility with the stress concentration arising around the pore periphery, under physiological loading conditions with respect to orthopedic hip implants. A comparative study of the two is performed with the aid of a newly defined optimizing parameter called pore efficiency that takes into consideration both the stiffness quantity and the stress localization around pores. To perform a detailed analysis of the response of the porous structure over the entire spectrum of loading conditions that a hip implant is subjected to in vivo, the mechanical responses of 3D finite element models of cubic and rectangular parallelepiped geometries, with porosities varying over a range of 10% to 60%, are simulated under representative compressive, flexural as well as combined loading conditions. The results that are obtained are used to suggest a range of pore diameters that lower the effective stiffness and modulus of the implant to around 60% of the stiffness and modulus of dense solid implants while keeping the stress levels within permissible limits.

No MeSH data available.


Related in: MedlinePlus