Pore Geometry Optimization of Titanium (Ti6Al4V) Alloy, for Its Application in the Fabrication of Customized Hip Implants.

Roy S, Panda D, Khutia N, Chowdhury AR - Int J Biomater (2014)

Bottom Line: The present study investigates the mechanical response of representative volume elements of porous Ti-6Al-4V alloy, to arrive at a desired range of pore geometries that would optimize the reduction in stiffness necessary for biocompatibility with the stress concentration arising around the pore periphery, under physiological loading conditions with respect to orthopedic hip implants.A comparative study of the two is performed with the aid of a newly defined optimizing parameter called pore efficiency that takes into consideration both the stiffness quantity and the stress localization around pores.The results that are obtained are used to suggest a range of pore diameters that lower the effective stiffness and modulus of the implant to around 60% of the stiffness and modulus of dense solid implants while keeping the stress levels within permissible limits.

View Article: PubMed Central - PubMed

Affiliation: Department of Aerospace Engineering and Applied Mechanics, Indian Institute of Engineering Science and Technology, Shibpur 711103, India.

ABSTRACT
The present study investigates the mechanical response of representative volume elements of porous Ti-6Al-4V alloy, to arrive at a desired range of pore geometries that would optimize the reduction in stiffness necessary for biocompatibility with the stress concentration arising around the pore periphery, under physiological loading conditions with respect to orthopedic hip implants. A comparative study of the two is performed with the aid of a newly defined optimizing parameter called pore efficiency that takes into consideration both the stiffness quantity and the stress localization around pores. To perform a detailed analysis of the response of the porous structure over the entire spectrum of loading conditions that a hip implant is subjected to in vivo, the mechanical responses of 3D finite element models of cubic and rectangular parallelepiped geometries, with porosities varying over a range of 10% to 60%, are simulated under representative compressive, flexural as well as combined loading conditions. The results that are obtained are used to suggest a range of pore diameters that lower the effective stiffness and modulus of the implant to around 60% of the stiffness and modulus of dense solid implants while keeping the stress levels within permissible limits.

No MeSH data available.


Related in: MedlinePlus

Plots depicting variation of pore efficiency under flexural loading, μf, with pore diameter, for different percentages of porosity for 3D porous FE beam models subjected to flexural loading only.
© Copyright Policy - open-access
Related In: Results  -  Collection


getmorefigures.php?uid=PMC4221907&req=5

fig13: Plots depicting variation of pore efficiency under flexural loading, μf, with pore diameter, for different percentages of porosity for 3D porous FE beam models subjected to flexural loading only.

Mentions: In order to optimize the effects of reduction in bending stiffness with the increasing stress concentration factor as aforesaid, a new parameter called pore efficiency for flexural loading, μf = 1/(σbKb,rel⁡), is introduced. Interestingly, it is found that the maximum values of pore efficiency that is attained for each of the different porosity percentages are in close proximity at a value of around 0.6 (Figure 13).This suggests that as we increase the porosity, the maximum values of relative bending stiffness and stress concentration factor decrease and increase, respectively, by equal proportion, as a result of which their net cumulative effect manages to remain more or less unchanged in terms of suitability. Thus, depending on the degree of stiffness reduction that is required to be achieved for perfect compatibility with the bone, a trade-off has to be made between the two conflicting parameters of relative bending stiffness and allowable stress concentration. Since stiffness reduction is the primary design criterion while designing our porous hip implant, a pore diameter in the range of 3.65 mm for 60% porosity is thus suggested as the optimum range of pore diameter. This pore diameter allows for up to 60% reduction in bending stiffness (Figure 11), while limiting the flexural stress concentration factor within 4.35 (Figure 12). Subsequently, from the investigation carried out by Bergmann et al. [19], it can be understood that a maximum von Mises stress of around 605 MPa is expected to be attained around the pore periphery near the neck of a 60% porous hip implant having the same shape as the designed solid hip implant under normal gait conditions for the above stress concentration of 4.35. On taking the yield strength value of annealed Ti-6Al-4V alloy under room temperature conditions to be around 895 MPa, an approximate factor of safety of 2.2 is thus obtained for the suggested value of pore diameter. Since while stumbling action, peak muscle forces as high as 2 times the normal hip joint contact forces might arise approximately [19], this factor of safety is found to be well on the conservative side.


Pore Geometry Optimization of Titanium (Ti6Al4V) Alloy, for Its Application in the Fabrication of Customized Hip Implants.

Roy S, Panda D, Khutia N, Chowdhury AR - Int J Biomater (2014)

Plots depicting variation of pore efficiency under flexural loading, μf, with pore diameter, for different percentages of porosity for 3D porous FE beam models subjected to flexural loading only.
© Copyright Policy - open-access
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC4221907&req=5

fig13: Plots depicting variation of pore efficiency under flexural loading, μf, with pore diameter, for different percentages of porosity for 3D porous FE beam models subjected to flexural loading only.
Mentions: In order to optimize the effects of reduction in bending stiffness with the increasing stress concentration factor as aforesaid, a new parameter called pore efficiency for flexural loading, μf = 1/(σbKb,rel⁡), is introduced. Interestingly, it is found that the maximum values of pore efficiency that is attained for each of the different porosity percentages are in close proximity at a value of around 0.6 (Figure 13).This suggests that as we increase the porosity, the maximum values of relative bending stiffness and stress concentration factor decrease and increase, respectively, by equal proportion, as a result of which their net cumulative effect manages to remain more or less unchanged in terms of suitability. Thus, depending on the degree of stiffness reduction that is required to be achieved for perfect compatibility with the bone, a trade-off has to be made between the two conflicting parameters of relative bending stiffness and allowable stress concentration. Since stiffness reduction is the primary design criterion while designing our porous hip implant, a pore diameter in the range of 3.65 mm for 60% porosity is thus suggested as the optimum range of pore diameter. This pore diameter allows for up to 60% reduction in bending stiffness (Figure 11), while limiting the flexural stress concentration factor within 4.35 (Figure 12). Subsequently, from the investigation carried out by Bergmann et al. [19], it can be understood that a maximum von Mises stress of around 605 MPa is expected to be attained around the pore periphery near the neck of a 60% porous hip implant having the same shape as the designed solid hip implant under normal gait conditions for the above stress concentration of 4.35. On taking the yield strength value of annealed Ti-6Al-4V alloy under room temperature conditions to be around 895 MPa, an approximate factor of safety of 2.2 is thus obtained for the suggested value of pore diameter. Since while stumbling action, peak muscle forces as high as 2 times the normal hip joint contact forces might arise approximately [19], this factor of safety is found to be well on the conservative side.

Bottom Line: The present study investigates the mechanical response of representative volume elements of porous Ti-6Al-4V alloy, to arrive at a desired range of pore geometries that would optimize the reduction in stiffness necessary for biocompatibility with the stress concentration arising around the pore periphery, under physiological loading conditions with respect to orthopedic hip implants.A comparative study of the two is performed with the aid of a newly defined optimizing parameter called pore efficiency that takes into consideration both the stiffness quantity and the stress localization around pores.The results that are obtained are used to suggest a range of pore diameters that lower the effective stiffness and modulus of the implant to around 60% of the stiffness and modulus of dense solid implants while keeping the stress levels within permissible limits.

View Article: PubMed Central - PubMed

Affiliation: Department of Aerospace Engineering and Applied Mechanics, Indian Institute of Engineering Science and Technology, Shibpur 711103, India.

ABSTRACT
The present study investigates the mechanical response of representative volume elements of porous Ti-6Al-4V alloy, to arrive at a desired range of pore geometries that would optimize the reduction in stiffness necessary for biocompatibility with the stress concentration arising around the pore periphery, under physiological loading conditions with respect to orthopedic hip implants. A comparative study of the two is performed with the aid of a newly defined optimizing parameter called pore efficiency that takes into consideration both the stiffness quantity and the stress localization around pores. To perform a detailed analysis of the response of the porous structure over the entire spectrum of loading conditions that a hip implant is subjected to in vivo, the mechanical responses of 3D finite element models of cubic and rectangular parallelepiped geometries, with porosities varying over a range of 10% to 60%, are simulated under representative compressive, flexural as well as combined loading conditions. The results that are obtained are used to suggest a range of pore diameters that lower the effective stiffness and modulus of the implant to around 60% of the stiffness and modulus of dense solid implants while keeping the stress levels within permissible limits.

No MeSH data available.


Related in: MedlinePlus