Fluid-dynamic optimal design of helical vascular graft for stenotic disturbed flow.

Ha H, Hwang D, Choi WR, Baek J, Lee SJ - PLoS ONE (2014)

Bottom Line: Results showed that the swirling intensity and helicity of the swirling flow have a linear relation with a modified Germano number (Gn*) of the helical pipe.In addition, the swirling flow generated a beneficial flow structure at the stenosis by reducing the size of the recirculation flow under steady and pulsatile flow conditions.Therefore, the beneficial effects of a helical graft on the flow field can be estimated by using the magnitude of Gn*.

View Article: PubMed Central - PubMed

Affiliation: Department of Mechanical Engineering, Pohang University of Science and Technology (POSTECH), Pohang, Gyeongbuk, Republic of Korea.

ABSTRACT
Although a helical configuration of a prosthetic vascular graft appears to be clinically beneficial in suppressing thrombosis and intimal hyperplasia, an optimization of a helical design has yet to be achieved because of the lack of a detailed understanding on hemodynamic features in helical grafts and their fluid dynamic influences. In the present study, the swirling flow in a helical graft was hypothesized to have beneficial influences on a disturbed flow structure such as stenotic flow. The characteristics of swirling flows generated by helical tubes with various helical pitches and curvatures were investigated to prove the hypothesis. The fluid dynamic influences of these helical tubes on stenotic flow were quantitatively analysed by using a particle image velocimetry technique. Results showed that the swirling intensity and helicity of the swirling flow have a linear relation with a modified Germano number (Gn*) of the helical pipe. In addition, the swirling flow generated a beneficial flow structure at the stenosis by reducing the size of the recirculation flow under steady and pulsatile flow conditions. Therefore, the beneficial effects of a helical graft on the flow field can be estimated by using the magnitude of Gn*. Finally, an optimized helical design with a maximum Gn* was suggested for the future design of a vascular graft.

Show MeSH

Related in: MedlinePlus

Schematic diagrams of the experimental set-up.(a) Flow circuit system, (b) PIV velocity field measurement system, (c) schematic of the stenosis model.
© Copyright Policy
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC4215892&req=5

pone-0111047-g003: Schematic diagrams of the experimental set-up.(a) Flow circuit system, (b) PIV velocity field measurement system, (c) schematic of the stenosis model.

Mentions: Figure 3a shows a schematic diagram of the flow circulating system used in this study. A total of 3 L working fluid was prepared in an acryl reservoir. A 15 W centrifugal pump circulated the working fluid through the circuit at a constant flow rate. A 0.5 L air container was installed as a fluidic low-pass filter using a three-way connector to stabilize possible fluctuations in flow rate [27], [28]. To measure the flow rate, a variable area-type flow meter (Visi-Float, Dwyer Instruments, Inc., Michigan, USA) was installed after proper re-calibration with the working fluid. The flow rate was controlled by the internal fluid valve of the flow meter.


Fluid-dynamic optimal design of helical vascular graft for stenotic disturbed flow.

Ha H, Hwang D, Choi WR, Baek J, Lee SJ - PLoS ONE (2014)

Schematic diagrams of the experimental set-up.(a) Flow circuit system, (b) PIV velocity field measurement system, (c) schematic of the stenosis model.
© Copyright Policy
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC4215892&req=5

pone-0111047-g003: Schematic diagrams of the experimental set-up.(a) Flow circuit system, (b) PIV velocity field measurement system, (c) schematic of the stenosis model.
Mentions: Figure 3a shows a schematic diagram of the flow circulating system used in this study. A total of 3 L working fluid was prepared in an acryl reservoir. A 15 W centrifugal pump circulated the working fluid through the circuit at a constant flow rate. A 0.5 L air container was installed as a fluidic low-pass filter using a three-way connector to stabilize possible fluctuations in flow rate [27], [28]. To measure the flow rate, a variable area-type flow meter (Visi-Float, Dwyer Instruments, Inc., Michigan, USA) was installed after proper re-calibration with the working fluid. The flow rate was controlled by the internal fluid valve of the flow meter.

Bottom Line: Results showed that the swirling intensity and helicity of the swirling flow have a linear relation with a modified Germano number (Gn*) of the helical pipe.In addition, the swirling flow generated a beneficial flow structure at the stenosis by reducing the size of the recirculation flow under steady and pulsatile flow conditions.Therefore, the beneficial effects of a helical graft on the flow field can be estimated by using the magnitude of Gn*.

View Article: PubMed Central - PubMed

Affiliation: Department of Mechanical Engineering, Pohang University of Science and Technology (POSTECH), Pohang, Gyeongbuk, Republic of Korea.

ABSTRACT
Although a helical configuration of a prosthetic vascular graft appears to be clinically beneficial in suppressing thrombosis and intimal hyperplasia, an optimization of a helical design has yet to be achieved because of the lack of a detailed understanding on hemodynamic features in helical grafts and their fluid dynamic influences. In the present study, the swirling flow in a helical graft was hypothesized to have beneficial influences on a disturbed flow structure such as stenotic flow. The characteristics of swirling flows generated by helical tubes with various helical pitches and curvatures were investigated to prove the hypothesis. The fluid dynamic influences of these helical tubes on stenotic flow were quantitatively analysed by using a particle image velocimetry technique. Results showed that the swirling intensity and helicity of the swirling flow have a linear relation with a modified Germano number (Gn*) of the helical pipe. In addition, the swirling flow generated a beneficial flow structure at the stenosis by reducing the size of the recirculation flow under steady and pulsatile flow conditions. Therefore, the beneficial effects of a helical graft on the flow field can be estimated by using the magnitude of Gn*. Finally, an optimized helical design with a maximum Gn* was suggested for the future design of a vascular graft.

Show MeSH
Related in: MedlinePlus