Limits...
Study of the effect of distance and misalignment between magnetically coupled coils for wireless power transfer in intraocular pressure measurement.

Rendon-Nava AE, Díaz-Méndez JA, Nino-de-Rivera L, Calleja-Arriaga W, Gil-Carrasco F, Díaz-Alonso D - ScientificWorldJournal (2014)

Bottom Line: Power transfer was done by magnetic induction coupling method, by placing one of the inductors of the Maxwell-Wien bridge circuit and the inductor of the implant in close proximity.The Maxwell-Wien bridge circuit was biased with a 10 MHz sinusoidal signal.In order to have a proper inductive coupling link, special care must be taken when placing the two coils in proximity to avoid misalignment between them.

View Article: PubMed Central - PubMed

Affiliation: Graduate Department, National Polytechnic Institute of Mexico (IPN), ESIME UPC, Avenida Santa Ana 1000, San Francisco Culhuacan, 04260 Mexico City, DF, Mexico.

ABSTRACT
An analysis of the effect of distance and alignment between two magnetically coupled coils for wireless power transfer in intraocular pressure measurement is presented. For measurement purposes, a system was fabricated consisting of an external device, which is a Maxwell-Wien bridge circuit variation, in charge of transferring energy to a biomedical implant and reading data from it. The biomedical implant is an RLC tank circuit, encapsulated by a polyimide coating. Power transfer was done by magnetic induction coupling method, by placing one of the inductors of the Maxwell-Wien bridge circuit and the inductor of the implant in close proximity. The Maxwell-Wien bridge circuit was biased with a 10 MHz sinusoidal signal. The analysis presented in this paper proves that wireless transmission of power for intraocular pressure measurement is feasible with the measurement system proposed. In order to have a proper inductive coupling link, special care must be taken when placing the two coils in proximity to avoid misalignment between them.

Show MeSH

Related in: MedlinePlus

Maxwell-Wien circuit variation proposed for energy and data transfer.
© Copyright Policy - open-access
Related In: Results  -  Collection


getmorefigures.php?uid=PMC4109074&req=5

fig5: Maxwell-Wien circuit variation proposed for energy and data transfer.

Mentions: The circuit proposed to receive data and to transfer power to the implant is a Maxwell-Wien bridge circuit variation [23], shown in Figure 5. Resistors R2 and R4 are of the same value to bring balance to the circuit. In the absence of the implant, that is, when the two coils Ls and Lx are far enough from each other, the Maxwell-Wien bridge will be balanced; in other words, the voltage between nodes B and C will be zero.


Study of the effect of distance and misalignment between magnetically coupled coils for wireless power transfer in intraocular pressure measurement.

Rendon-Nava AE, Díaz-Méndez JA, Nino-de-Rivera L, Calleja-Arriaga W, Gil-Carrasco F, Díaz-Alonso D - ScientificWorldJournal (2014)

Maxwell-Wien circuit variation proposed for energy and data transfer.
© Copyright Policy - open-access
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC4109074&req=5

fig5: Maxwell-Wien circuit variation proposed for energy and data transfer.
Mentions: The circuit proposed to receive data and to transfer power to the implant is a Maxwell-Wien bridge circuit variation [23], shown in Figure 5. Resistors R2 and R4 are of the same value to bring balance to the circuit. In the absence of the implant, that is, when the two coils Ls and Lx are far enough from each other, the Maxwell-Wien bridge will be balanced; in other words, the voltage between nodes B and C will be zero.

Bottom Line: Power transfer was done by magnetic induction coupling method, by placing one of the inductors of the Maxwell-Wien bridge circuit and the inductor of the implant in close proximity.The Maxwell-Wien bridge circuit was biased with a 10 MHz sinusoidal signal.In order to have a proper inductive coupling link, special care must be taken when placing the two coils in proximity to avoid misalignment between them.

View Article: PubMed Central - PubMed

Affiliation: Graduate Department, National Polytechnic Institute of Mexico (IPN), ESIME UPC, Avenida Santa Ana 1000, San Francisco Culhuacan, 04260 Mexico City, DF, Mexico.

ABSTRACT
An analysis of the effect of distance and alignment between two magnetically coupled coils for wireless power transfer in intraocular pressure measurement is presented. For measurement purposes, a system was fabricated consisting of an external device, which is a Maxwell-Wien bridge circuit variation, in charge of transferring energy to a biomedical implant and reading data from it. The biomedical implant is an RLC tank circuit, encapsulated by a polyimide coating. Power transfer was done by magnetic induction coupling method, by placing one of the inductors of the Maxwell-Wien bridge circuit and the inductor of the implant in close proximity. The Maxwell-Wien bridge circuit was biased with a 10 MHz sinusoidal signal. The analysis presented in this paper proves that wireless transmission of power for intraocular pressure measurement is feasible with the measurement system proposed. In order to have a proper inductive coupling link, special care must be taken when placing the two coils in proximity to avoid misalignment between them.

Show MeSH
Related in: MedlinePlus