Numerically optimized radiofrequency pulses for robust and low-power cardiovascular T 2 preparation at 3T

View Article: PubMed Central - HTML

No MeSH data available.


Performance of the tailored pulses. a) Example characteristics of the HS1-derived TR-FOCI pulse. Note that the optimized pulse is broad compared to the standard HS1 (dotted line). b) Slice-excitation profile of the same HS1-derived TR-FOCI pulse. Above a low pulse power of γB1≈200 Hz, the profile has a very sharp transition. c) On-resonance inversion power requirements of the standard and TR-FOCI pulses. Once the HS1 pulse is optimized as a TR-FOCI pulse, it requires power similar to the standard HS8 pulse. d) Robustness of the same pulses to B0 variability at pulse power γB1 = 500 Hz. While the poor original profile of the HS8 pulse is improved after optimization, it is especially the optimized HS1 that has improved robustness.
© Copyright Policy - open-access
Related In: Results  -  Collection

License 1 - License 2
getmorefigures.php?uid=PMC4043734&req=5

Figure 1: Performance of the tailored pulses. a) Example characteristics of the HS1-derived TR-FOCI pulse. Note that the optimized pulse is broad compared to the standard HS1 (dotted line). b) Slice-excitation profile of the same HS1-derived TR-FOCI pulse. Above a low pulse power of γB1≈200 Hz, the profile has a very sharp transition. c) On-resonance inversion power requirements of the standard and TR-FOCI pulses. Once the HS1 pulse is optimized as a TR-FOCI pulse, it requires power similar to the standard HS8 pulse. d) Robustness of the same pulses to B0 variability at pulse power γB1 = 500 Hz. While the poor original profile of the HS8 pulse is improved after optimization, it is especially the optimized HS1 that has improved robustness.

Mentions: The optimized pulses demonstrated superior performance in the simulations compared to standard HSn pulses (Figure 1). The TR-FOCI pulses required 54% less power than the HS1 pulse to achieve artifact-free images and stable CNR (Figure 2), while images obtained with an HS8 pulse were never artifact-free. The optimized pulses needed roughly half the energy of the standard pulses, and the entire pulse sequence resulted in 20% less overall SAR deposition in the volunteers for artifact-free images with similar CNR as the original images.


Numerically optimized radiofrequency pulses for robust and low-power cardiovascular T 2 preparation at 3T
Performance of the tailored pulses. a) Example characteristics of the HS1-derived TR-FOCI pulse. Note that the optimized pulse is broad compared to the standard HS1 (dotted line). b) Slice-excitation profile of the same HS1-derived TR-FOCI pulse. Above a low pulse power of γB1≈200 Hz, the profile has a very sharp transition. c) On-resonance inversion power requirements of the standard and TR-FOCI pulses. Once the HS1 pulse is optimized as a TR-FOCI pulse, it requires power similar to the standard HS8 pulse. d) Robustness of the same pulses to B0 variability at pulse power γB1 = 500 Hz. While the poor original profile of the HS8 pulse is improved after optimization, it is especially the optimized HS1 that has improved robustness.
© Copyright Policy - open-access
Related In: Results  -  Collection

License 1 - License 2
Show All Figures
getmorefigures.php?uid=PMC4043734&req=5

Figure 1: Performance of the tailored pulses. a) Example characteristics of the HS1-derived TR-FOCI pulse. Note that the optimized pulse is broad compared to the standard HS1 (dotted line). b) Slice-excitation profile of the same HS1-derived TR-FOCI pulse. Above a low pulse power of γB1≈200 Hz, the profile has a very sharp transition. c) On-resonance inversion power requirements of the standard and TR-FOCI pulses. Once the HS1 pulse is optimized as a TR-FOCI pulse, it requires power similar to the standard HS8 pulse. d) Robustness of the same pulses to B0 variability at pulse power γB1 = 500 Hz. While the poor original profile of the HS8 pulse is improved after optimization, it is especially the optimized HS1 that has improved robustness.
Mentions: The optimized pulses demonstrated superior performance in the simulations compared to standard HSn pulses (Figure 1). The TR-FOCI pulses required 54% less power than the HS1 pulse to achieve artifact-free images and stable CNR (Figure 2), while images obtained with an HS8 pulse were never artifact-free. The optimized pulses needed roughly half the energy of the standard pulses, and the entire pulse sequence resulted in 20% less overall SAR deposition in the volunteers for artifact-free images with similar CNR as the original images.

View Article: PubMed Central - HTML

No MeSH data available.