A special phase detector for magnetic inductive measurement of cerebral hemorrhage.

Jin G, Sun J, Qin M - PLoS ONE (2014)

Bottom Line: The noise and drift decreased as the frequency decreased.The results are in agreement with those from previous reports.The results from the injection group showed a similar trend of increasing phase shift change with increasing injection volume.

View Article: PubMed Central - PubMed

Affiliation: College of Biomedical Engineering, Third Military Medical University, Chongqing, China.

ABSTRACT
Cerebral hemorrhage is an important clinical problem that is often monitored and studied with expensive techniques, such as computed tomography (CT), magnetic resonance imaging (MRI), and positron emission tomography (PET). These devices are not readily available in economically underdeveloped regions of the world and in emergency departments and emergency zones. The magnetic inductive method is an emerging technology that may become a new tool to detect cerebral hemorrhage. In this study, a special phase detector (PD) was developed and used for cerebral hemorrhage detection with the magnetic inductive method. The performance indicated that the PD can achieve phase noise as low as 6 m° and a 4-hour phase drift as low as 30 m° at 21.4 MHz. The noise and drift decreased as the frequency decreased. The performance at 10.7 MHz was slightly better than that of other recently developed phase detection systems. To test the practicality of the system, the PD was used to detect the volume change in a self-made physical model of the brain. The measured phase shift was approximately proportional to the volume change of physiological saline inside the model. The change of the phase shift increased as the volume change and frequency increased. The results are in agreement with those from previous reports. To verify the feasibility of in vivo detection, an autologous blood injection model was established in rabbit brain. The results from the injection group showed a similar trend of increasing phase shift change with increasing injection volume. The average phase shift change induced by a 3-ml injection of blood was 0.502°±0.119°, which was much larger than that of the control group. The measurement system can distinguish a minimal cerebral hemorrhage volume of approximately 0.5 ml. All of the results demonstrated that the PD used with this method can detect cerebral hemorrhage.

Show MeSH

Related in: MedlinePlus

Phase drifts of the physical hemorrhage model over 4 hours without injection and pumping.
© Copyright Policy
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC4016262&req=5

pone-0097179-g006: Phase drifts of the physical hemorrhage model over 4 hours without injection and pumping.

Mentions: Figure 6 shows the phase shifts for the physical model over 4 hours at the three frequencies. The data were normalized to the average value. The phase drifts were 27 m°, 44 m° and 87 m° at frequencies of 1 MHz, 10.7 MHz and 21.4 MHz, respectively.


A special phase detector for magnetic inductive measurement of cerebral hemorrhage.

Jin G, Sun J, Qin M - PLoS ONE (2014)

Phase drifts of the physical hemorrhage model over 4 hours without injection and pumping.
© Copyright Policy
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC4016262&req=5

pone-0097179-g006: Phase drifts of the physical hemorrhage model over 4 hours without injection and pumping.
Mentions: Figure 6 shows the phase shifts for the physical model over 4 hours at the three frequencies. The data were normalized to the average value. The phase drifts were 27 m°, 44 m° and 87 m° at frequencies of 1 MHz, 10.7 MHz and 21.4 MHz, respectively.

Bottom Line: The noise and drift decreased as the frequency decreased.The results are in agreement with those from previous reports.The results from the injection group showed a similar trend of increasing phase shift change with increasing injection volume.

View Article: PubMed Central - PubMed

Affiliation: College of Biomedical Engineering, Third Military Medical University, Chongqing, China.

ABSTRACT
Cerebral hemorrhage is an important clinical problem that is often monitored and studied with expensive techniques, such as computed tomography (CT), magnetic resonance imaging (MRI), and positron emission tomography (PET). These devices are not readily available in economically underdeveloped regions of the world and in emergency departments and emergency zones. The magnetic inductive method is an emerging technology that may become a new tool to detect cerebral hemorrhage. In this study, a special phase detector (PD) was developed and used for cerebral hemorrhage detection with the magnetic inductive method. The performance indicated that the PD can achieve phase noise as low as 6 m° and a 4-hour phase drift as low as 30 m° at 21.4 MHz. The noise and drift decreased as the frequency decreased. The performance at 10.7 MHz was slightly better than that of other recently developed phase detection systems. To test the practicality of the system, the PD was used to detect the volume change in a self-made physical model of the brain. The measured phase shift was approximately proportional to the volume change of physiological saline inside the model. The change of the phase shift increased as the volume change and frequency increased. The results are in agreement with those from previous reports. To verify the feasibility of in vivo detection, an autologous blood injection model was established in rabbit brain. The results from the injection group showed a similar trend of increasing phase shift change with increasing injection volume. The average phase shift change induced by a 3-ml injection of blood was 0.502°±0.119°, which was much larger than that of the control group. The measurement system can distinguish a minimal cerebral hemorrhage volume of approximately 0.5 ml. All of the results demonstrated that the PD used with this method can detect cerebral hemorrhage.

Show MeSH
Related in: MedlinePlus