Limits...
The reduction methods of operator's radiation dose for portable dental X-ray machines.

Cho JY, Han WJ - Restor Dent Endod (2012)

Bottom Line: Two kinds of portable dental X-ray machines (DX3000, Dexcowin and Rextar, Posdion) were used.And the backscatter shield reduced operator's radiation dose at the operator's chest and waist levels to 0.1 - 37%.When portable dental X-ray systems are used, it is recommended to select X-ray machine attached with a backscatter shield and a long cone and to wear the lead gloves.

View Article: PubMed Central - PubMed

Affiliation: Department of Oral & Maxillofacial Radiology, Dankook University College of Dentistry, Cheonan, Korea.

ABSTRACT

Objectives: This study was aimed to investigate the methods to reduce operator's radiation dose when taking intraoral radiographs with portable dental X-ray machines.

Materials and methods: Two kinds of portable dental X-ray machines (DX3000, Dexcowin and Rextar, Posdion) were used. Operator's radiation dose was measured with an 1,800 cc ionization chamber (RadCal Corp.) at the hand level of X-ray tubehead and at the operator's chest and waist levels with and without the backscatter shield. The operator's radiation dose at the hand level was measured with and without lead gloves and with long and short cones.

Results: The backscatter shield reduced operator's radiation dose at the hand level of X-ray tubehead to 23 - 32%, the lead gloves to 26 - 31%, and long cone to 48 - 52%. And the backscatter shield reduced operator's radiation dose at the operator's chest and waist levels to 0.1 - 37%.

Conclusions: When portable dental X-ray systems are used, it is recommended to select X-ray machine attached with a backscatter shield and a long cone and to wear the lead gloves.

No MeSH data available.


Operator's radiation dose was measured by an ionization chamber at the operator's hand level using the portable dental X-ray machine without a lead shield.
© Copyright Policy - open-access
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC3569401&req=5

Figure 4: Operator's radiation dose was measured by an ionization chamber at the operator's hand level using the portable dental X-ray machine without a lead shield.

Mentions: Dose measurement was carried out at the operator's hand level during the exposure of lower anterior and posterior teeth, with and without a lead backscatter shield attached at the end of cone. The setting for exposure was 60 kVp, 1 mA and 0.4 seconds for the lower anterior teeth, 0.8 seconds for the lower posterior teeth using digital sensor (Figures 4 and 5).


The reduction methods of operator's radiation dose for portable dental X-ray machines.

Cho JY, Han WJ - Restor Dent Endod (2012)

Operator's radiation dose was measured by an ionization chamber at the operator's hand level using the portable dental X-ray machine without a lead shield.
© Copyright Policy - open-access
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC3569401&req=5

Figure 4: Operator's radiation dose was measured by an ionization chamber at the operator's hand level using the portable dental X-ray machine without a lead shield.
Mentions: Dose measurement was carried out at the operator's hand level during the exposure of lower anterior and posterior teeth, with and without a lead backscatter shield attached at the end of cone. The setting for exposure was 60 kVp, 1 mA and 0.4 seconds for the lower anterior teeth, 0.8 seconds for the lower posterior teeth using digital sensor (Figures 4 and 5).

Bottom Line: Two kinds of portable dental X-ray machines (DX3000, Dexcowin and Rextar, Posdion) were used.And the backscatter shield reduced operator's radiation dose at the operator's chest and waist levels to 0.1 - 37%.When portable dental X-ray systems are used, it is recommended to select X-ray machine attached with a backscatter shield and a long cone and to wear the lead gloves.

View Article: PubMed Central - PubMed

Affiliation: Department of Oral & Maxillofacial Radiology, Dankook University College of Dentistry, Cheonan, Korea.

ABSTRACT

Objectives: This study was aimed to investigate the methods to reduce operator's radiation dose when taking intraoral radiographs with portable dental X-ray machines.

Materials and methods: Two kinds of portable dental X-ray machines (DX3000, Dexcowin and Rextar, Posdion) were used. Operator's radiation dose was measured with an 1,800 cc ionization chamber (RadCal Corp.) at the hand level of X-ray tubehead and at the operator's chest and waist levels with and without the backscatter shield. The operator's radiation dose at the hand level was measured with and without lead gloves and with long and short cones.

Results: The backscatter shield reduced operator's radiation dose at the hand level of X-ray tubehead to 23 - 32%, the lead gloves to 26 - 31%, and long cone to 48 - 52%. And the backscatter shield reduced operator's radiation dose at the operator's chest and waist levels to 0.1 - 37%.

Conclusions: When portable dental X-ray systems are used, it is recommended to select X-ray machine attached with a backscatter shield and a long cone and to wear the lead gloves.

No MeSH data available.