Dosimetric characterizations of GZP6 (60)Co high dose rate brachytherapy sources: application of superimposition method.

Bahreyni Toossi MT, Ghorbani M, Mowlavi AA, Meigooni AS - Radiol Oncol (2012)

Bottom Line: These parameters of this source are compared with the available data for Ralstron (60)Co and microSelectron(192)Ir sources.The results of these investigations show that the dosimetric parameters of GZP6 source are comparable to those for the Ralstron source.The calculated dosimetric quantities of this new source can be introduced as input data to the GZP6 treatment planning system (TPS) and to validate the performance of the TPS.

View Article: PubMed Central - PubMed

Affiliation: Medical Physics Research Center, Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran.

ABSTRACT

Background: Dosimetric characteristics of a high dose rate (HDR) GZP6 Co-60 brachytherapy source have been evaluated following American Association of Physicists in MedicineTask Group 43U1 (AAPM TG-43U1) recommendations for their clinical applications.

Materials and methods: MCNP-4C and MCNPX Monte Carlo codes were utilized to calculate dose rate constant, two dimensional (2D) dose distribution, radial dose function and 2D anisotropy function of the source. These parameters of this source are compared with the available data for Ralstron (60)Co and microSelectron(192)Ir sources. Besides, a superimposition method was developed to extend the obtained results for the GZP6 source No. 3 to other GZP6 sources.

Results: The simulated value for dose rate constant for GZP6 source was 1.104±0.03 cGyh-1U-1. The graphical and tabulated radial dose function and 2D anisotropy function of this source are presented here. The results of these investigations show that the dosimetric parameters of GZP6 source are comparable to those for the Ralstron source. While dose rate constant for the two (60)Co sources are similar to that for the microSelectron(192)Ir source, there are differences between radial dose function and anisotropy functions. Radial dose function of the (192)Ir source is less steep than both (60)Co source models. In addition, the (60)Co sources are showing more isotropic dose distribution than the (192)Ir source.

Conclusions: The superimposition method is applicable to produce dose distributions for other source arrangements from the dose distribution of a single source. The calculated dosimetric quantities of this new source can be introduced as input data to the GZP6 treatment planning system (TPS) and to validate the performance of the TPS.

No MeSH data available.


Dose distributions (Gy) obtained by MC simulations (blue lines) and superimposition method (red lines) for GZP6 sources: (a) source No. 1; (b) source No. 2 and (c) source No. 5. Isodoses of 1–20 Gy are contoured in the figure and since the contours by the two methods are overlapped in many points, are almost not distinguishable.
© Copyright Policy - open-access
Related In: Results  -  Collection

License 1 - License 2
getmorefigures.php?uid=PMC3472933&req=5

f5-rado-46-02-170: Dose distributions (Gy) obtained by MC simulations (blue lines) and superimposition method (red lines) for GZP6 sources: (a) source No. 1; (b) source No. 2 and (c) source No. 5. Isodoses of 1–20 Gy are contoured in the figure and since the contours by the two methods are overlapped in many points, are almost not distinguishable.

Mentions: The dose distributions for the GZP6 source No. 1, 2 and 5 which was obtained by MC simulations of the sources and the superimposition method described are presented in the Figure 5. The dose contours of 1, 2, 5, 10, 15 and 20 Gy were plotted. As it can be observed the dose distributions obtained by the two methods for the sources 1, 2 and 5 are equal.


Dosimetric characterizations of GZP6 (60)Co high dose rate brachytherapy sources: application of superimposition method.

Bahreyni Toossi MT, Ghorbani M, Mowlavi AA, Meigooni AS - Radiol Oncol (2012)

Dose distributions (Gy) obtained by MC simulations (blue lines) and superimposition method (red lines) for GZP6 sources: (a) source No. 1; (b) source No. 2 and (c) source No. 5. Isodoses of 1–20 Gy are contoured in the figure and since the contours by the two methods are overlapped in many points, are almost not distinguishable.
© Copyright Policy - open-access
Related In: Results  -  Collection

License 1 - License 2
Show All Figures
getmorefigures.php?uid=PMC3472933&req=5

f5-rado-46-02-170: Dose distributions (Gy) obtained by MC simulations (blue lines) and superimposition method (red lines) for GZP6 sources: (a) source No. 1; (b) source No. 2 and (c) source No. 5. Isodoses of 1–20 Gy are contoured in the figure and since the contours by the two methods are overlapped in many points, are almost not distinguishable.
Mentions: The dose distributions for the GZP6 source No. 1, 2 and 5 which was obtained by MC simulations of the sources and the superimposition method described are presented in the Figure 5. The dose contours of 1, 2, 5, 10, 15 and 20 Gy were plotted. As it can be observed the dose distributions obtained by the two methods for the sources 1, 2 and 5 are equal.

Bottom Line: These parameters of this source are compared with the available data for Ralstron (60)Co and microSelectron(192)Ir sources.The results of these investigations show that the dosimetric parameters of GZP6 source are comparable to those for the Ralstron source.The calculated dosimetric quantities of this new source can be introduced as input data to the GZP6 treatment planning system (TPS) and to validate the performance of the TPS.

View Article: PubMed Central - PubMed

Affiliation: Medical Physics Research Center, Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran.

ABSTRACT

Background: Dosimetric characteristics of a high dose rate (HDR) GZP6 Co-60 brachytherapy source have been evaluated following American Association of Physicists in MedicineTask Group 43U1 (AAPM TG-43U1) recommendations for their clinical applications.

Materials and methods: MCNP-4C and MCNPX Monte Carlo codes were utilized to calculate dose rate constant, two dimensional (2D) dose distribution, radial dose function and 2D anisotropy function of the source. These parameters of this source are compared with the available data for Ralstron (60)Co and microSelectron(192)Ir sources. Besides, a superimposition method was developed to extend the obtained results for the GZP6 source No. 3 to other GZP6 sources.

Results: The simulated value for dose rate constant for GZP6 source was 1.104±0.03 cGyh-1U-1. The graphical and tabulated radial dose function and 2D anisotropy function of this source are presented here. The results of these investigations show that the dosimetric parameters of GZP6 source are comparable to those for the Ralstron source. While dose rate constant for the two (60)Co sources are similar to that for the microSelectron(192)Ir source, there are differences between radial dose function and anisotropy functions. Radial dose function of the (192)Ir source is less steep than both (60)Co source models. In addition, the (60)Co sources are showing more isotropic dose distribution than the (192)Ir source.

Conclusions: The superimposition method is applicable to produce dose distributions for other source arrangements from the dose distribution of a single source. The calculated dosimetric quantities of this new source can be introduced as input data to the GZP6 treatment planning system (TPS) and to validate the performance of the TPS.

No MeSH data available.