Limits...
A wireless magnetic resonance energy transfer system for micro implantable medical sensors.

Li X, Zhang H, Peng F, Li Y, Yang T, Wang B, Fang D - Sensors (Basel) (2012)

Bottom Line: The energy transfer efficiency of the four-coil system is greatly improved compared to the conventional two-coil system.In addition, the output current varies with changes in the distance.The whole implanted part is packaged with PDMS of excellent biocompatibility and the volume of it is about 1 cm(3).

View Article: PubMed Central - PubMed

Affiliation: School of Electronics and Information Engineering, Beijing Jiaotong University, Beijing 100044, China. lixiuhan@bjtu.edu.cn

ABSTRACT
Based on the magnetic resonance coupling principle, in this paper a wireless energy transfer system is designed and implemented for the power supply of micro-implantable medical sensors. The entire system is composed of the in vitro part, including the energy transmitting circuit and resonant transmitter coils, and in vivo part, including the micro resonant receiver coils and signal shaping chip which includes the rectifier module and LDO voltage regulator module. Transmitter and receiver coils are wound by Litz wire, and the diameter of the receiver coils is just 1.9 cm. The energy transfer efficiency of the four-coil system is greatly improved compared to the conventional two-coil system. When the distance between the transmitter coils and the receiver coils is 1.5 cm, the transfer efficiency is 85% at the frequency of 742 kHz. The power transfer efficiency can be optimized by adding magnetic enhanced resonators. The receiving voltage signal is converted to a stable output voltage of 3.3 V and a current of 10 mA at the distance of 2 cm. In addition, the output current varies with changes in the distance. The whole implanted part is packaged with PDMS of excellent biocompatibility and the volume of it is about 1 cm(3).

Show MeSH

Related in: MedlinePlus

The experimental setup for the four-coil power transfer system.
© Copyright Policy
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC3472828&req=5

f10-sensors-12-10292: The experimental setup for the four-coil power transfer system.

Mentions: As shown in Figure 10, a network analyzer (Agilent 8714ET) is used to measure the electrical specifications of the coils and the transfer efficiency of the wireless power transfer system. The coils' electrical specifications i.e., L, Q, dc resistance, and ac resistance can be measured through the Smith Chart with the frequency sweeping from 300 kHz to 10 MHz. The power transfer efficiency can be obtained from the S-parameter:(19)f(dB)=10×Lg[S212/(1−S112)]


A wireless magnetic resonance energy transfer system for micro implantable medical sensors.

Li X, Zhang H, Peng F, Li Y, Yang T, Wang B, Fang D - Sensors (Basel) (2012)

The experimental setup for the four-coil power transfer system.
© Copyright Policy
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC3472828&req=5

f10-sensors-12-10292: The experimental setup for the four-coil power transfer system.
Mentions: As shown in Figure 10, a network analyzer (Agilent 8714ET) is used to measure the electrical specifications of the coils and the transfer efficiency of the wireless power transfer system. The coils' electrical specifications i.e., L, Q, dc resistance, and ac resistance can be measured through the Smith Chart with the frequency sweeping from 300 kHz to 10 MHz. The power transfer efficiency can be obtained from the S-parameter:(19)f(dB)=10×Lg[S212/(1−S112)]

Bottom Line: The energy transfer efficiency of the four-coil system is greatly improved compared to the conventional two-coil system.In addition, the output current varies with changes in the distance.The whole implanted part is packaged with PDMS of excellent biocompatibility and the volume of it is about 1 cm(3).

View Article: PubMed Central - PubMed

Affiliation: School of Electronics and Information Engineering, Beijing Jiaotong University, Beijing 100044, China. lixiuhan@bjtu.edu.cn

ABSTRACT
Based on the magnetic resonance coupling principle, in this paper a wireless energy transfer system is designed and implemented for the power supply of micro-implantable medical sensors. The entire system is composed of the in vitro part, including the energy transmitting circuit and resonant transmitter coils, and in vivo part, including the micro resonant receiver coils and signal shaping chip which includes the rectifier module and LDO voltage regulator module. Transmitter and receiver coils are wound by Litz wire, and the diameter of the receiver coils is just 1.9 cm. The energy transfer efficiency of the four-coil system is greatly improved compared to the conventional two-coil system. When the distance between the transmitter coils and the receiver coils is 1.5 cm, the transfer efficiency is 85% at the frequency of 742 kHz. The power transfer efficiency can be optimized by adding magnetic enhanced resonators. The receiving voltage signal is converted to a stable output voltage of 3.3 V and a current of 10 mA at the distance of 2 cm. In addition, the output current varies with changes in the distance. The whole implanted part is packaged with PDMS of excellent biocompatibility and the volume of it is about 1 cm(3).

Show MeSH
Related in: MedlinePlus