Clinical use of aided cortical auditory evoked potentials as a measure of physiological detection or physiological discrimination.

Billings CJ, Papesh MA, Penman TM, Baltzell LS, Gallun FJ - Int J Otolaryngol (2012)

Bottom Line: One major contributor to this ambiguity is the wide range of variability across published studies and across individuals within a given study; some results demonstrate expected amplification effects, while others demonstrate limited or no amplification effects.Recent evidence indicates that some of the variability in amplification effects may be explained by distinguishing between experiments that focused on physiological detection of a stimulus versus those that differentiate responses to two audible signals, or physiological discrimination.Stimulus levels were varied to study the effect of hearing-aid-signal/hearing-aid-noise audibility relative to the noise-masked thresholds.

View Article: PubMed Central - PubMed

Affiliation: National Center for Rehabilitative Auditory Research, Portland Veterans Affairs Medical Center, Portland, OR 97239, USA ; Department of Otolaryngology/Head & Neck Surgery, Oregon Health & Science University, Portland, OR 97239, USA.

ABSTRACT
The clinical usefulness of aided cortical auditory evoked potentials (CAEPs) remains unclear despite several decades of research. One major contributor to this ambiguity is the wide range of variability across published studies and across individuals within a given study; some results demonstrate expected amplification effects, while others demonstrate limited or no amplification effects. Recent evidence indicates that some of the variability in amplification effects may be explained by distinguishing between experiments that focused on physiological detection of a stimulus versus those that differentiate responses to two audible signals, or physiological discrimination. Herein, we ask if either of these approaches is clinically feasible given the inherent challenges with aided CAEPs. N1 and P2 waves were elicited from 12 noise-masked normal-hearing individuals using hearing-aid-processed 1000-Hz pure tones. Stimulus levels were varied to study the effect of hearing-aid-signal/hearing-aid-noise audibility relative to the noise-masked thresholds. Results demonstrate that clinical use of aided CAEPs may be justified when determining whether audible stimuli are physiologically detectable relative to inaudible signals. However, differentiating aided CAEPs elicited from two suprathreshold stimuli (i.e., physiological discrimination) is problematic and should not be used for clinical decision making until a better understanding of the interaction between hearing-aid-processed stimuli and CAEPs can be established.

No MeSH data available.


Related in: MedlinePlus

Experimental design. One-third octave band levels at 1000 Hz are shown for the three hearing-aid-processed recordings. Scaling of the recordings resulted in Near θ, Low, Mid, and High conditions. The shaded background shows the background noise masker level relative to hearing aid signal and noise levels.
© Copyright Policy - open-access
Related In: Results  -  Collection


getmorefigures.php?uid=PMC3472537&req=5

fig2: Experimental design. One-third octave band levels at 1000 Hz are shown for the three hearing-aid-processed recordings. Scaling of the recordings resulted in Near θ, Low, Mid, and High conditions. The shaded background shows the background noise masker level relative to hearing aid signal and noise levels.

Mentions: A schematic depicting the hearing aid signal (i.e., 1000-Hz tone) and noise levels in relation to the background noise masker for each hearing aid condition is shown in Figure 2. For low-input recordings from Hearing Aid A and Hearing Aid B, four scaling factors were calculated that adjusted the level of the tone and hearing aid noise in approximately 10 dB steps to produce the following presentation levels: Near θ, a near-threshold level at which both tone and hearing aid noise were inaudible due to the noise masker amplitude (intended to represent an inaudible unaided condition); Low, a level at which the signal was above the level of the noise masker but the hearing aid noise was below; Mid and High, two levels at which the signal and hearing aid noise were both audible but absolute signal level differed. Due to the higher ratio of signal level to hearing aid noise obtained in the higher input level recording from Hearing Aid B, five stimulus levels were necessary in order to span the range of audible and inaudible signals. In total, 13 experimental conditions were presented to each participant. Notice that, within each block of hearing aid conditions, the SNR between the signal level and the hearing aid noise remains constant (i.e., approx. 11 dB for Hearing Aid A (Recording 1), 8 dB for Hearing Aid B (Recording 2), 23 dB for Hearing Aid B (Recording 3), while the overall stimulus output level changes in 10 dB steps. However, the effective SNR was much smaller for Near θ and Low conditions due to the background noise masker. Levels displayed in Figure 2 are 1/3 octave band values centered at 1000 Hz measured with a Brüel and Kjær 2260 Investigator sound level meter fitted with a Brüel & Kjær ear simulator (number 4157).


Clinical use of aided cortical auditory evoked potentials as a measure of physiological detection or physiological discrimination.

Billings CJ, Papesh MA, Penman TM, Baltzell LS, Gallun FJ - Int J Otolaryngol (2012)

Experimental design. One-third octave band levels at 1000 Hz are shown for the three hearing-aid-processed recordings. Scaling of the recordings resulted in Near θ, Low, Mid, and High conditions. The shaded background shows the background noise masker level relative to hearing aid signal and noise levels.
© Copyright Policy - open-access
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC3472537&req=5

fig2: Experimental design. One-third octave band levels at 1000 Hz are shown for the three hearing-aid-processed recordings. Scaling of the recordings resulted in Near θ, Low, Mid, and High conditions. The shaded background shows the background noise masker level relative to hearing aid signal and noise levels.
Mentions: A schematic depicting the hearing aid signal (i.e., 1000-Hz tone) and noise levels in relation to the background noise masker for each hearing aid condition is shown in Figure 2. For low-input recordings from Hearing Aid A and Hearing Aid B, four scaling factors were calculated that adjusted the level of the tone and hearing aid noise in approximately 10 dB steps to produce the following presentation levels: Near θ, a near-threshold level at which both tone and hearing aid noise were inaudible due to the noise masker amplitude (intended to represent an inaudible unaided condition); Low, a level at which the signal was above the level of the noise masker but the hearing aid noise was below; Mid and High, two levels at which the signal and hearing aid noise were both audible but absolute signal level differed. Due to the higher ratio of signal level to hearing aid noise obtained in the higher input level recording from Hearing Aid B, five stimulus levels were necessary in order to span the range of audible and inaudible signals. In total, 13 experimental conditions were presented to each participant. Notice that, within each block of hearing aid conditions, the SNR between the signal level and the hearing aid noise remains constant (i.e., approx. 11 dB for Hearing Aid A (Recording 1), 8 dB for Hearing Aid B (Recording 2), 23 dB for Hearing Aid B (Recording 3), while the overall stimulus output level changes in 10 dB steps. However, the effective SNR was much smaller for Near θ and Low conditions due to the background noise masker. Levels displayed in Figure 2 are 1/3 octave band values centered at 1000 Hz measured with a Brüel and Kjær 2260 Investigator sound level meter fitted with a Brüel & Kjær ear simulator (number 4157).

Bottom Line: One major contributor to this ambiguity is the wide range of variability across published studies and across individuals within a given study; some results demonstrate expected amplification effects, while others demonstrate limited or no amplification effects.Recent evidence indicates that some of the variability in amplification effects may be explained by distinguishing between experiments that focused on physiological detection of a stimulus versus those that differentiate responses to two audible signals, or physiological discrimination.Stimulus levels were varied to study the effect of hearing-aid-signal/hearing-aid-noise audibility relative to the noise-masked thresholds.

View Article: PubMed Central - PubMed

Affiliation: National Center for Rehabilitative Auditory Research, Portland Veterans Affairs Medical Center, Portland, OR 97239, USA ; Department of Otolaryngology/Head & Neck Surgery, Oregon Health & Science University, Portland, OR 97239, USA.

ABSTRACT
The clinical usefulness of aided cortical auditory evoked potentials (CAEPs) remains unclear despite several decades of research. One major contributor to this ambiguity is the wide range of variability across published studies and across individuals within a given study; some results demonstrate expected amplification effects, while others demonstrate limited or no amplification effects. Recent evidence indicates that some of the variability in amplification effects may be explained by distinguishing between experiments that focused on physiological detection of a stimulus versus those that differentiate responses to two audible signals, or physiological discrimination. Herein, we ask if either of these approaches is clinically feasible given the inherent challenges with aided CAEPs. N1 and P2 waves were elicited from 12 noise-masked normal-hearing individuals using hearing-aid-processed 1000-Hz pure tones. Stimulus levels were varied to study the effect of hearing-aid-signal/hearing-aid-noise audibility relative to the noise-masked thresholds. Results demonstrate that clinical use of aided CAEPs may be justified when determining whether audible stimuli are physiologically detectable relative to inaudible signals. However, differentiating aided CAEPs elicited from two suprathreshold stimuli (i.e., physiological discrimination) is problematic and should not be used for clinical decision making until a better understanding of the interaction between hearing-aid-processed stimuli and CAEPs can be established.

No MeSH data available.


Related in: MedlinePlus