A parametric study delineating irreversible electroporation from thermal damage based on a minimally invasive intracranial procedure.

Garcia PA, Rossmeisl JH, Neal RE, Ellis TL, Davalos RV - Biomed Eng Online (2011)

Bottom Line: We developed numerical simulations of typical protocols based on a previously published computed tomographic (CT) guided in vivo procedure.We confirm that determining an IRE treatment protocol requires incorporating all the physical effects of electroporation, and that these effects may have significant implications in treatment planning and outcome assessment.The goal of the manuscript is to provide the reader with the numerical methods to assess multiple-pulse electroporation treatment protocols in order to isolate IRE from thermal damage and capitalize on the benefits of a non-thermal mode of tissue ablation.

View Article: PubMed Central - HTML - PubMed

Affiliation: Bioelectromechanical Systems Laboratory, School of Biomedical Engineering and Sciences, Virginia Tech-Wake Forest University, Blacksburg, VA, USA.

ABSTRACT

Background: Irreversible electroporation (IRE) is a new minimally invasive technique to kill undesirable tissue in a non-thermal manner. In order to maximize the benefits from an IRE procedure, the pulse parameters and electrode configuration must be optimized to achieve complete coverage of the targeted tissue while preventing thermal damage due to excessive Joule heating.

Methods: We developed numerical simulations of typical protocols based on a previously published computed tomographic (CT) guided in vivo procedure. These models were adapted to assess the effects of temperature, electroporation, pulse duration, and repetition rate on the volumes of tissue undergoing IRE alone or in superposition with thermal damage.

Results: Nine different combinations of voltage and pulse frequency were investigated, five of which resulted in IRE alone while four produced IRE in superposition with thermal damage.

Conclusions: The parametric study evaluated the influence of pulse frequency and applied voltage on treatment volumes, and refined a proposed method to delineate IRE from thermal damage. We confirm that determining an IRE treatment protocol requires incorporating all the physical effects of electroporation, and that these effects may have significant implications in treatment planning and outcome assessment. The goal of the manuscript is to provide the reader with the numerical methods to assess multiple-pulse electroporation treatment protocols in order to isolate IRE from thermal damage and capitalize on the benefits of a non-thermal mode of tissue ablation.

Show MeSH

Related in: MedlinePlus

Time history of the total current during an eighty pulse (50 μs) IRE treatment. The pulses were delivered at frequencies of A) 0.5 Hz (160 s), B) 1 Hz (80 s), and C) 4 Hz (20 s). The applied voltages were 500 V, 1000 V, and 1500 V for each frequency investigated.
© Copyright Policy - open-access
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC3108916&req=5

Figure 7: Time history of the total current during an eighty pulse (50 μs) IRE treatment. The pulses were delivered at frequencies of A) 0.5 Hz (160 s), B) 1 Hz (80 s), and C) 4 Hz (20 s). The applied voltages were 500 V, 1000 V, and 1500 V for each frequency investigated.

Mentions: In addition to monitoring the temperature during the experimental procedure, the current of each individual pulse was measured by the NanoKnife®. It was found that the current throughout the procedure was 1.11 ± 0.2 A [23]. The resulting currents from the parametric IRE simulations for the 500, 1000, and 1500 V treatments delivered at 0.5, 1, and 4 Hz were calculated and are displayed in Figure 7. Applying 500 V resulted in electrical currents of 1.08 - 1.12 A independent of the pulse repetition rate. A current ranging between 2.53 - 2.98 A was calculated when using 1000 V. Simulations at 1500 V resulted in an initial current of 3.96 A, and reached 4.33 A (0.5 Hz), 4.66 A (1 Hz), or 5.90 A (4 Hz) at the completion of the pulses. The variation between the calculated currents can be explained by the increase in temperature during the pulse delivery. When the pulses were delivered at 4 Hz, there was less time for the heat to dissipate through conduction or blood perfusion. Therefore, the calculated temperatures were significantly higher which also resulted in higher electrical conductivity and thus electrical current. The increase in electrical current was not observed during the 500 V treatments since at this lower voltage the thermal effects are negligible compared to the changes in conductivity due to electroporation. The measured current agreed with the calculated currents, validating our assumption of an increase in the brain electric conductivity by a factor of 3.0 due to electroporation.


A parametric study delineating irreversible electroporation from thermal damage based on a minimally invasive intracranial procedure.

Garcia PA, Rossmeisl JH, Neal RE, Ellis TL, Davalos RV - Biomed Eng Online (2011)

Time history of the total current during an eighty pulse (50 μs) IRE treatment. The pulses were delivered at frequencies of A) 0.5 Hz (160 s), B) 1 Hz (80 s), and C) 4 Hz (20 s). The applied voltages were 500 V, 1000 V, and 1500 V for each frequency investigated.
© Copyright Policy - open-access
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC3108916&req=5

Figure 7: Time history of the total current during an eighty pulse (50 μs) IRE treatment. The pulses were delivered at frequencies of A) 0.5 Hz (160 s), B) 1 Hz (80 s), and C) 4 Hz (20 s). The applied voltages were 500 V, 1000 V, and 1500 V for each frequency investigated.
Mentions: In addition to monitoring the temperature during the experimental procedure, the current of each individual pulse was measured by the NanoKnife®. It was found that the current throughout the procedure was 1.11 ± 0.2 A [23]. The resulting currents from the parametric IRE simulations for the 500, 1000, and 1500 V treatments delivered at 0.5, 1, and 4 Hz were calculated and are displayed in Figure 7. Applying 500 V resulted in electrical currents of 1.08 - 1.12 A independent of the pulse repetition rate. A current ranging between 2.53 - 2.98 A was calculated when using 1000 V. Simulations at 1500 V resulted in an initial current of 3.96 A, and reached 4.33 A (0.5 Hz), 4.66 A (1 Hz), or 5.90 A (4 Hz) at the completion of the pulses. The variation between the calculated currents can be explained by the increase in temperature during the pulse delivery. When the pulses were delivered at 4 Hz, there was less time for the heat to dissipate through conduction or blood perfusion. Therefore, the calculated temperatures were significantly higher which also resulted in higher electrical conductivity and thus electrical current. The increase in electrical current was not observed during the 500 V treatments since at this lower voltage the thermal effects are negligible compared to the changes in conductivity due to electroporation. The measured current agreed with the calculated currents, validating our assumption of an increase in the brain electric conductivity by a factor of 3.0 due to electroporation.

Bottom Line: We developed numerical simulations of typical protocols based on a previously published computed tomographic (CT) guided in vivo procedure.We confirm that determining an IRE treatment protocol requires incorporating all the physical effects of electroporation, and that these effects may have significant implications in treatment planning and outcome assessment.The goal of the manuscript is to provide the reader with the numerical methods to assess multiple-pulse electroporation treatment protocols in order to isolate IRE from thermal damage and capitalize on the benefits of a non-thermal mode of tissue ablation.

View Article: PubMed Central - HTML - PubMed

Affiliation: Bioelectromechanical Systems Laboratory, School of Biomedical Engineering and Sciences, Virginia Tech-Wake Forest University, Blacksburg, VA, USA.

ABSTRACT

Background: Irreversible electroporation (IRE) is a new minimally invasive technique to kill undesirable tissue in a non-thermal manner. In order to maximize the benefits from an IRE procedure, the pulse parameters and electrode configuration must be optimized to achieve complete coverage of the targeted tissue while preventing thermal damage due to excessive Joule heating.

Methods: We developed numerical simulations of typical protocols based on a previously published computed tomographic (CT) guided in vivo procedure. These models were adapted to assess the effects of temperature, electroporation, pulse duration, and repetition rate on the volumes of tissue undergoing IRE alone or in superposition with thermal damage.

Results: Nine different combinations of voltage and pulse frequency were investigated, five of which resulted in IRE alone while four produced IRE in superposition with thermal damage.

Conclusions: The parametric study evaluated the influence of pulse frequency and applied voltage on treatment volumes, and refined a proposed method to delineate IRE from thermal damage. We confirm that determining an IRE treatment protocol requires incorporating all the physical effects of electroporation, and that these effects may have significant implications in treatment planning and outcome assessment. The goal of the manuscript is to provide the reader with the numerical methods to assess multiple-pulse electroporation treatment protocols in order to isolate IRE from thermal damage and capitalize on the benefits of a non-thermal mode of tissue ablation.