Theory and validation of magnetic resonance fluid motion estimation using intensity flow data.

Wong KK, Kelso RM, Worthley SG, Sanders P, Mazumdar J, Abbott D - PLoS ONE (2009)

Bottom Line: The analytical and computed flow fields are compared to provide an error estimate of our methodology.The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding.This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

View Article: PubMed Central - PubMed

Affiliation: Centre for Biomedical Engineering and School of Electrical & Electronic Engineering, University of Adelaide, Adelaide, South Australia, Australia. kelvin.wong@adelaide.edu.au

ABSTRACT

Background: Motion tracking based on spatial-temporal radio-frequency signals from the pixel representation of magnetic resonance (MR) imaging of a non-stationary fluid is able to provide two dimensional vector field maps. This supports the underlying fundamentals of magnetic resonance fluid motion estimation and generates a new methodology for flow measurement that is based on registration of nuclear signals from moving hydrogen nuclei in fluid. However, there is a need to validate the computational aspect of the approach by using velocity flow field data that we will assume as the true reference information or ground truth.

Methodology/principal findings: In this study, we create flow vectors based on an ideal analytical vortex, and generate artificial signal-motion image data to verify our computational approach. The analytical and computed flow fields are compared to provide an error estimate of our methodology. The comparison shows that the fluid motion estimation approach using simulated MR data is accurate and robust enough for flow field mapping. To verify our methodology, we have tested the computational configuration on magnetic resonance images of cardiac blood and proved that the theory of magnetic resonance fluid motion estimation can be applicable practically.

Conclusions/significance: The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding. This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

Show MeSH

Related in: MedlinePlus

Tracking accuracy of rotation based on variation of noise in image.Comparison of analytical and computational flow velocities in the radial direction is illustrated. Velocity profiles and their differences are shown in (a,d,g,j) and (b,e,h,k) respectively. Quadrant of the vortical grid is displayed in (c,f,i,l). Variation of feature density adjusts the moderation and extent of flow grid prediction.
© Copyright Policy
Related In: Results  -  Collection


getmorefigures.php?uid=PMC2651647&req=5

pone-0004747-g009: Tracking accuracy of rotation based on variation of noise in image.Comparison of analytical and computational flow velocities in the radial direction is illustrated. Velocity profiles and their differences are shown in (a,d,g,j) and (b,e,h,k) respectively. Quadrant of the vortical grid is displayed in (c,f,i,l). Variation of feature density adjusts the moderation and extent of flow grid prediction.

Mentions: We have applied different smoothing filter masks that have widths at 1 by 1, 5 by 5 and 9 by 9 pixels. Gaussian noise is added to the images at percentage (%) of 0, 10, 30 and 50 incrementally. Note that the overall average error increases for radius ranging from 1 to 5 mm for every stage of additional noise input into the tracking images. However, when smoothing is applied to the images, the velocity profile improves in accuracy with respect to the analytical one as a result of the suppression of noise signals. This can be illustrated by parts (a) to (h) of Figure 9, that corresponds to velocity and error profiles derived from tracking based on 0, 10, and 30% noise to image addition. Nevertheless, we also observed that for higher noise added to images with percentages such as 30% or 50%, over smoothing using a (9×9) pixel filter mask size reduces the tracking accuracy relative to that when a (5×5) pixel smoothing mask is applied. The results describe the effect of over-smoothing after the noise addition has reached a certain threshold.


Theory and validation of magnetic resonance fluid motion estimation using intensity flow data.

Wong KK, Kelso RM, Worthley SG, Sanders P, Mazumdar J, Abbott D - PLoS ONE (2009)

Tracking accuracy of rotation based on variation of noise in image.Comparison of analytical and computational flow velocities in the radial direction is illustrated. Velocity profiles and their differences are shown in (a,d,g,j) and (b,e,h,k) respectively. Quadrant of the vortical grid is displayed in (c,f,i,l). Variation of feature density adjusts the moderation and extent of flow grid prediction.
© Copyright Policy
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC2651647&req=5

pone-0004747-g009: Tracking accuracy of rotation based on variation of noise in image.Comparison of analytical and computational flow velocities in the radial direction is illustrated. Velocity profiles and their differences are shown in (a,d,g,j) and (b,e,h,k) respectively. Quadrant of the vortical grid is displayed in (c,f,i,l). Variation of feature density adjusts the moderation and extent of flow grid prediction.
Mentions: We have applied different smoothing filter masks that have widths at 1 by 1, 5 by 5 and 9 by 9 pixels. Gaussian noise is added to the images at percentage (%) of 0, 10, 30 and 50 incrementally. Note that the overall average error increases for radius ranging from 1 to 5 mm for every stage of additional noise input into the tracking images. However, when smoothing is applied to the images, the velocity profile improves in accuracy with respect to the analytical one as a result of the suppression of noise signals. This can be illustrated by parts (a) to (h) of Figure 9, that corresponds to velocity and error profiles derived from tracking based on 0, 10, and 30% noise to image addition. Nevertheless, we also observed that for higher noise added to images with percentages such as 30% or 50%, over smoothing using a (9×9) pixel filter mask size reduces the tracking accuracy relative to that when a (5×5) pixel smoothing mask is applied. The results describe the effect of over-smoothing after the noise addition has reached a certain threshold.

Bottom Line: The analytical and computed flow fields are compared to provide an error estimate of our methodology.The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding.This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

View Article: PubMed Central - PubMed

Affiliation: Centre for Biomedical Engineering and School of Electrical & Electronic Engineering, University of Adelaide, Adelaide, South Australia, Australia. kelvin.wong@adelaide.edu.au

ABSTRACT

Background: Motion tracking based on spatial-temporal radio-frequency signals from the pixel representation of magnetic resonance (MR) imaging of a non-stationary fluid is able to provide two dimensional vector field maps. This supports the underlying fundamentals of magnetic resonance fluid motion estimation and generates a new methodology for flow measurement that is based on registration of nuclear signals from moving hydrogen nuclei in fluid. However, there is a need to validate the computational aspect of the approach by using velocity flow field data that we will assume as the true reference information or ground truth.

Methodology/principal findings: In this study, we create flow vectors based on an ideal analytical vortex, and generate artificial signal-motion image data to verify our computational approach. The analytical and computed flow fields are compared to provide an error estimate of our methodology. The comparison shows that the fluid motion estimation approach using simulated MR data is accurate and robust enough for flow field mapping. To verify our methodology, we have tested the computational configuration on magnetic resonance images of cardiac blood and proved that the theory of magnetic resonance fluid motion estimation can be applicable practically.

Conclusions/significance: The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding. This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.