Theory and validation of magnetic resonance fluid motion estimation using intensity flow data.

Wong KK, Kelso RM, Worthley SG, Sanders P, Mazumdar J, Abbott D - PLoS ONE (2009)

Bottom Line: The analytical and computed flow fields are compared to provide an error estimate of our methodology.The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding.This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

View Article: PubMed Central - PubMed

Affiliation: Centre for Biomedical Engineering and School of Electrical & Electronic Engineering, University of Adelaide, Adelaide, South Australia, Australia. kelvin.wong@adelaide.edu.au

ABSTRACT

Background: Motion tracking based on spatial-temporal radio-frequency signals from the pixel representation of magnetic resonance (MR) imaging of a non-stationary fluid is able to provide two dimensional vector field maps. This supports the underlying fundamentals of magnetic resonance fluid motion estimation and generates a new methodology for flow measurement that is based on registration of nuclear signals from moving hydrogen nuclei in fluid. However, there is a need to validate the computational aspect of the approach by using velocity flow field data that we will assume as the true reference information or ground truth.

Methodology/principal findings: In this study, we create flow vectors based on an ideal analytical vortex, and generate artificial signal-motion image data to verify our computational approach. The analytical and computed flow fields are compared to provide an error estimate of our methodology. The comparison shows that the fluid motion estimation approach using simulated MR data is accurate and robust enough for flow field mapping. To verify our methodology, we have tested the computational configuration on magnetic resonance images of cardiac blood and proved that the theory of magnetic resonance fluid motion estimation can be applicable practically.

Conclusions/significance: The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding. This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

Show MeSH

Related in: MedlinePlus

Tracking accuracy based on angular velocities.Flow vector angular differences are shown based on variations of sampling window size versus image size with different levels of track interval dimensions (i.e. levels 1 to 3). Similar to the results shown for variation of angular velocity, this set of graphs demonstrates that the tracking is more stable for sampling window sizes from a recommendation of 11 pixels onwards. Angular velocity error decreases as the image becomes dimensionally bigger.
© Copyright Policy
Related In: Results  -  Collection


getmorefigures.php?uid=PMC2651647&req=5

pone-0004747-g007: Tracking accuracy based on angular velocities.Flow vector angular differences are shown based on variations of sampling window size versus image size with different levels of track interval dimensions (i.e. levels 1 to 3). Similar to the results shown for variation of angular velocity, this set of graphs demonstrates that the tracking is more stable for sampling window sizes from a recommendation of 11 pixels onwards. Angular velocity error decreases as the image becomes dimensionally bigger.

Mentions: Larger image sizes give a reduction in error estimates and can be further reduced by a corresponding increment in the sampling window size used by the algorithm for image sizes from 260 by 260 pixels onwards (Figure 7). The improvement in tracking accuracy due to increase in image size can be explained by the enhancement of feature quality due to the larger number of pixels used to represent varying space of high and low intensity track intervals. In addition to this observation and explanation, we can further deduce that an increase of sampling window size at the right image proportion will improve tracking accuracy due to capture of the required features at appropriate interrogation space.


Theory and validation of magnetic resonance fluid motion estimation using intensity flow data.

Wong KK, Kelso RM, Worthley SG, Sanders P, Mazumdar J, Abbott D - PLoS ONE (2009)

Tracking accuracy based on angular velocities.Flow vector angular differences are shown based on variations of sampling window size versus image size with different levels of track interval dimensions (i.e. levels 1 to 3). Similar to the results shown for variation of angular velocity, this set of graphs demonstrates that the tracking is more stable for sampling window sizes from a recommendation of 11 pixels onwards. Angular velocity error decreases as the image becomes dimensionally bigger.
© Copyright Policy
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC2651647&req=5

pone-0004747-g007: Tracking accuracy based on angular velocities.Flow vector angular differences are shown based on variations of sampling window size versus image size with different levels of track interval dimensions (i.e. levels 1 to 3). Similar to the results shown for variation of angular velocity, this set of graphs demonstrates that the tracking is more stable for sampling window sizes from a recommendation of 11 pixels onwards. Angular velocity error decreases as the image becomes dimensionally bigger.
Mentions: Larger image sizes give a reduction in error estimates and can be further reduced by a corresponding increment in the sampling window size used by the algorithm for image sizes from 260 by 260 pixels onwards (Figure 7). The improvement in tracking accuracy due to increase in image size can be explained by the enhancement of feature quality due to the larger number of pixels used to represent varying space of high and low intensity track intervals. In addition to this observation and explanation, we can further deduce that an increase of sampling window size at the right image proportion will improve tracking accuracy due to capture of the required features at appropriate interrogation space.

Bottom Line: The analytical and computed flow fields are compared to provide an error estimate of our methodology.The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding.This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

View Article: PubMed Central - PubMed

Affiliation: Centre for Biomedical Engineering and School of Electrical & Electronic Engineering, University of Adelaide, Adelaide, South Australia, Australia. kelvin.wong@adelaide.edu.au

ABSTRACT

Background: Motion tracking based on spatial-temporal radio-frequency signals from the pixel representation of magnetic resonance (MR) imaging of a non-stationary fluid is able to provide two dimensional vector field maps. This supports the underlying fundamentals of magnetic resonance fluid motion estimation and generates a new methodology for flow measurement that is based on registration of nuclear signals from moving hydrogen nuclei in fluid. However, there is a need to validate the computational aspect of the approach by using velocity flow field data that we will assume as the true reference information or ground truth.

Methodology/principal findings: In this study, we create flow vectors based on an ideal analytical vortex, and generate artificial signal-motion image data to verify our computational approach. The analytical and computed flow fields are compared to provide an error estimate of our methodology. The comparison shows that the fluid motion estimation approach using simulated MR data is accurate and robust enough for flow field mapping. To verify our methodology, we have tested the computational configuration on magnetic resonance images of cardiac blood and proved that the theory of magnetic resonance fluid motion estimation can be applicable practically.

Conclusions/significance: The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding. This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.