Theory and validation of magnetic resonance fluid motion estimation using intensity flow data.

Wong KK, Kelso RM, Worthley SG, Sanders P, Mazumdar J, Abbott D - PLoS ONE (2009)

Bottom Line: The analytical and computed flow fields are compared to provide an error estimate of our methodology.The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding.This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

View Article: PubMed Central - PubMed

Affiliation: Centre for Biomedical Engineering and School of Electrical & Electronic Engineering, University of Adelaide, Adelaide, South Australia, Australia. kelvin.wong@adelaide.edu.au

ABSTRACT

Background: Motion tracking based on spatial-temporal radio-frequency signals from the pixel representation of magnetic resonance (MR) imaging of a non-stationary fluid is able to provide two dimensional vector field maps. This supports the underlying fundamentals of magnetic resonance fluid motion estimation and generates a new methodology for flow measurement that is based on registration of nuclear signals from moving hydrogen nuclei in fluid. However, there is a need to validate the computational aspect of the approach by using velocity flow field data that we will assume as the true reference information or ground truth.

Methodology/principal findings: In this study, we create flow vectors based on an ideal analytical vortex, and generate artificial signal-motion image data to verify our computational approach. The analytical and computed flow fields are compared to provide an error estimate of our methodology. The comparison shows that the fluid motion estimation approach using simulated MR data is accurate and robust enough for flow field mapping. To verify our methodology, we have tested the computational configuration on magnetic resonance images of cardiac blood and proved that the theory of magnetic resonance fluid motion estimation can be applicable practically.

Conclusions/significance: The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding. This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

Show MeSH

Related in: MedlinePlus

Multi-resolution motion estimation using pyramid implementation.Diagrammatic view of the Gaussian pyramid with optical flow applied onto every image level (0 to L) is presented in (a). Each level in the pyramid is a sub-sampled version of the level below. In the first step, the optical flow between the top level images is computed. We project the computed coarse-level flow field onto the next finer pyramid level and continue this at each level of the pyramid until the finest pyramid level has been reached. The system block diagram in (b) gives an illustration of the algorithmic operation of this pyramidal implementation.
© Copyright Policy
Related In: Results  -  Collection


getmorefigures.php?uid=PMC2651647&req=5

pone-0004747-g003: Multi-resolution motion estimation using pyramid implementation.Diagrammatic view of the Gaussian pyramid with optical flow applied onto every image level (0 to L) is presented in (a). Each level in the pyramid is a sub-sampled version of the level below. In the first step, the optical flow between the top level images is computed. We project the computed coarse-level flow field onto the next finer pyramid level and continue this at each level of the pyramid until the finest pyramid level has been reached. The system block diagram in (b) gives an illustration of the algorithmic operation of this pyramidal implementation.

Mentions: In our optical flow scheme, the pyramidal Lucas Kanade optical flow method [7], which incorporates a multi-scale approach, has been applied to support large scale fluid motion and for improved accuracy. A top-down estimation of the flow by using an image pyramid is performed, with the apex representing the MR image at a coarse scale. Computational results from this level are passed to the next and this process is carried on based on the flow estimated at the preceding scale until the original scale is reached. We refer to the diagram in Figure 3 to illustrate the computational aspect of pyramidal optical flow. Projection of the computed coarse-level flow field onto the next finer pyramid level is continued for each level of the pyramid until the finest pyramid level has been reached.


Theory and validation of magnetic resonance fluid motion estimation using intensity flow data.

Wong KK, Kelso RM, Worthley SG, Sanders P, Mazumdar J, Abbott D - PLoS ONE (2009)

Multi-resolution motion estimation using pyramid implementation.Diagrammatic view of the Gaussian pyramid with optical flow applied onto every image level (0 to L) is presented in (a). Each level in the pyramid is a sub-sampled version of the level below. In the first step, the optical flow between the top level images is computed. We project the computed coarse-level flow field onto the next finer pyramid level and continue this at each level of the pyramid until the finest pyramid level has been reached. The system block diagram in (b) gives an illustration of the algorithmic operation of this pyramidal implementation.
© Copyright Policy
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC2651647&req=5

pone-0004747-g003: Multi-resolution motion estimation using pyramid implementation.Diagrammatic view of the Gaussian pyramid with optical flow applied onto every image level (0 to L) is presented in (a). Each level in the pyramid is a sub-sampled version of the level below. In the first step, the optical flow between the top level images is computed. We project the computed coarse-level flow field onto the next finer pyramid level and continue this at each level of the pyramid until the finest pyramid level has been reached. The system block diagram in (b) gives an illustration of the algorithmic operation of this pyramidal implementation.
Mentions: In our optical flow scheme, the pyramidal Lucas Kanade optical flow method [7], which incorporates a multi-scale approach, has been applied to support large scale fluid motion and for improved accuracy. A top-down estimation of the flow by using an image pyramid is performed, with the apex representing the MR image at a coarse scale. Computational results from this level are passed to the next and this process is carried on based on the flow estimated at the preceding scale until the original scale is reached. We refer to the diagram in Figure 3 to illustrate the computational aspect of pyramidal optical flow. Projection of the computed coarse-level flow field onto the next finer pyramid level is continued for each level of the pyramid until the finest pyramid level has been reached.

Bottom Line: The analytical and computed flow fields are compared to provide an error estimate of our methodology.The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding.This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

View Article: PubMed Central - PubMed

Affiliation: Centre for Biomedical Engineering and School of Electrical & Electronic Engineering, University of Adelaide, Adelaide, South Australia, Australia. kelvin.wong@adelaide.edu.au

ABSTRACT

Background: Motion tracking based on spatial-temporal radio-frequency signals from the pixel representation of magnetic resonance (MR) imaging of a non-stationary fluid is able to provide two dimensional vector field maps. This supports the underlying fundamentals of magnetic resonance fluid motion estimation and generates a new methodology for flow measurement that is based on registration of nuclear signals from moving hydrogen nuclei in fluid. However, there is a need to validate the computational aspect of the approach by using velocity flow field data that we will assume as the true reference information or ground truth.

Methodology/principal findings: In this study, we create flow vectors based on an ideal analytical vortex, and generate artificial signal-motion image data to verify our computational approach. The analytical and computed flow fields are compared to provide an error estimate of our methodology. The comparison shows that the fluid motion estimation approach using simulated MR data is accurate and robust enough for flow field mapping. To verify our methodology, we have tested the computational configuration on magnetic resonance images of cardiac blood and proved that the theory of magnetic resonance fluid motion estimation can be applicable practically.

Conclusions/significance: The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding. This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.