Robotic cadaver testing of a new total ankle prosthesis model (GERMAN ANKLE SYSTEM)

View Article: PubMed Central - HTML

AUTOMATICALLY GENERATED EXCERPT
Please rate it.

An investigation should be carried out into possible increased forces, torques and altered motions during motion of a load-bearing ankle after implantation of two different total ankle prostheses... A robot-based cadaver test was developed for the study... We included two different ankle prostheses (Hintegra, Newdeal SA, Vienne, France; German Ankle System, Small Bone Innovations, Morrisville, PA, USA)... The robot detected the load-bearing (30 kg) motion of the specimens without prostheses during 100 cycles and mimicked that exact motion after the prostheses were implanted for another 100 cycles... The resulting forces and torques were recorded by an integrated load cell (model FT Delta SI-660-60; Schunk, Lauffen, Germany)... The spatial orientation of the tibia, fibula, and foot plate was recorded via an ultrasound measurement system (model CMS HS; Zebris Inc., Tuebingen, Germany)... The Hintegra and German Ankle System, significantly increased the forces and torques in relation to the specimen without prosthesis with one exception [One-sample-t-test, each p ≤ 0.01 (exception, parameter lateral force measured with the German Ankle System, p = 0.34)]... The force, torque and motion differences between the specimens before and after implantation of the prostheses were lower with the German Ankle System than with the Hintegra (unpaired t-test, each p ≤ 0.05)... In conclusion, the German Ankle System prosthesis had less of an affect on resulting forces and torques during partial-weight bearing ankle motion than the Hintegra prosthesis... This might improve function and minimize loosening during the clinical use.

No MeSH data available.


Setting with robot, specimen, and motion analysis system. Specimen mounted to the robot and footplate and equipped with ultrasound transducers.
© Copyright Policy
Related In: Results  -  Collection


getmorefigures.php?uid=PMC2562089&req=5

Figure 1: Setting with robot, specimen, and motion analysis system. Specimen mounted to the robot and footplate and equipped with ultrasound transducers.

Mentions: We included two different ankle prostheses (Hintegra, Newdeal SA, Vienne, France; German Ankle System, Small Bone Innovations, Morrisville, PA, USA). The prostheses were implanted in seven paired cadaver specimens (sides and testing sequence randomized). The specimens were mounted on an industrial robot (RX 90, Stäubli Tec-Systems, Bayreuth, Germany, Figure 1).


Robotic cadaver testing of a new total ankle prosthesis model (GERMAN ANKLE SYSTEM)
Setting with robot, specimen, and motion analysis system. Specimen mounted to the robot and footplate and equipped with ultrasound transducers.
© Copyright Policy
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC2562089&req=5

Figure 1: Setting with robot, specimen, and motion analysis system. Specimen mounted to the robot and footplate and equipped with ultrasound transducers.
Mentions: We included two different ankle prostheses (Hintegra, Newdeal SA, Vienne, France; German Ankle System, Small Bone Innovations, Morrisville, PA, USA). The prostheses were implanted in seven paired cadaver specimens (sides and testing sequence randomized). The specimens were mounted on an industrial robot (RX 90, Stäubli Tec-Systems, Bayreuth, Germany, Figure 1).

View Article: PubMed Central - HTML

AUTOMATICALLY GENERATED EXCERPT
Please rate it.

An investigation should be carried out into possible increased forces, torques and altered motions during motion of a load-bearing ankle after implantation of two different total ankle prostheses... A robot-based cadaver test was developed for the study... We included two different ankle prostheses (Hintegra, Newdeal SA, Vienne, France; German Ankle System, Small Bone Innovations, Morrisville, PA, USA)... The robot detected the load-bearing (30 kg) motion of the specimens without prostheses during 100 cycles and mimicked that exact motion after the prostheses were implanted for another 100 cycles... The resulting forces and torques were recorded by an integrated load cell (model FT Delta SI-660-60; Schunk, Lauffen, Germany)... The spatial orientation of the tibia, fibula, and foot plate was recorded via an ultrasound measurement system (model CMS HS; Zebris Inc., Tuebingen, Germany)... The Hintegra and German Ankle System, significantly increased the forces and torques in relation to the specimen without prosthesis with one exception [One-sample-t-test, each p ≤ 0.01 (exception, parameter lateral force measured with the German Ankle System, p = 0.34)]... The force, torque and motion differences between the specimens before and after implantation of the prostheses were lower with the German Ankle System than with the Hintegra (unpaired t-test, each p ≤ 0.05)... In conclusion, the German Ankle System prosthesis had less of an affect on resulting forces and torques during partial-weight bearing ankle motion than the Hintegra prosthesis... This might improve function and minimize loosening during the clinical use.