Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications.

Heine JJ, Thomas JA - Biomed Eng Online (2008)

Bottom Line: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared.Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically.Calibration data collection can be reduced by applying a simple normalization technique.

View Article: PubMed Central - HTML - PubMed

Affiliation: Cancer Prevention & Control Department, Moffitt Cancer Center, Tampa, Florida, USA. john.heine@moffitt.org

ABSTRACT

Background: Breast density is a significant breast cancer risk factor. Currently, there is no standard method for measuring this important factor. Work presented here represents an essential component of an ongoing project that seeks to determine the appropriate method for calibrating (standardizing) mammography image data to account for the x-ray image acquisition influences. Longer term goals of this project are to make accurate breast density measurements in support of risk studies.

Methods: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared. Normalization methods were studied to assess the possibility of reducing the amount of calibration data collection. The percent glandular calibration map functional form was investigated. Spatial variations in the calibration data were used to assess the uncertainty in the calibration application by applying error propagation analyses.

Results: Logarithmic response curves are well approximated as linear. Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically. Calibration data collection can be reduced by applying a simple normalization technique. The calibration map is well approximated as linear. Intrasystem calibration variation was on the order of four percent, which was approximately half of the intersystem variation.

Conclusion: FFDM systems provide a quantitative output, and the calibration quantities presented here may be used for data acquired on similar FFDM systems.

Show MeSH

Related in: MedlinePlus

Glandular tissue logarithmic response sector example plots. These plots show the example sector response curves for the Florida (diamonds) and Kansas (asterisks) locations for the 26 kVp – Mo/Mo settings. The fitted regression line is solid for Florida and dashed for Kansas.
© Copyright Policy - open-access
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC2365951&req=5

Figure 3: Glandular tissue logarithmic response sector example plots. These plots show the example sector response curves for the Florida (diamonds) and Kansas (asterisks) locations for the 26 kVp – Mo/Mo settings. The fitted regression line is solid for Florida and dashed for Kansas.

Mentions: Variation in R2 is not provided because it was very close to zero in all cases. Individual sector LR example plots for the Mo/Mo – 26 kVp settings are shown in Fig. 3 (glandular) and Fig. 4 (fatty). Plots for a given tissue type are shown in the same figure for intersystem comparisons. In these plots, measured points from the FL location are denoted by diamonds and asterisks for the KS measured points. Fitted lines from the regression analysis are solid for the FL plots and dashed for KS plots. Regression parameters are listed in Table 2. Regression parameter distribution quantities for the other Mo/Rh and Rh/Rh settings are shown in Table 3. The corresponding individual sector plot examples for these other settings are not shown because they were similar to the prior examples. For all distribution examples, variation within given location is relatively small, and the effective attenuation coefficients are in close agreement between the locations. Moreover, the linear model approximation holds when assessing the R2 statistic for all situations. To show the intersystem agreement, a paired t-test was applied by pairing all of the 10 average attenuation coefficient quantities for the FL location with the corresponding values from the KS location (second column from Table 1 and Table 3), which gave p = 0.56 (linear correlation of = 0.97). Following the same procedure for the 10 average intercept quantities (using the fifth columns from the same tables), gave p = 0.101 (linear correlation = 0.99). Using the p value as a gauge, effective attenuation coefficients as an ensemble are similar across the two systems. Intersystem intercept comparisons indicate a weak system dependence.


Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications.

Heine JJ, Thomas JA - Biomed Eng Online (2008)

Glandular tissue logarithmic response sector example plots. These plots show the example sector response curves for the Florida (diamonds) and Kansas (asterisks) locations for the 26 kVp – Mo/Mo settings. The fitted regression line is solid for Florida and dashed for Kansas.
© Copyright Policy - open-access
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC2365951&req=5

Figure 3: Glandular tissue logarithmic response sector example plots. These plots show the example sector response curves for the Florida (diamonds) and Kansas (asterisks) locations for the 26 kVp – Mo/Mo settings. The fitted regression line is solid for Florida and dashed for Kansas.
Mentions: Variation in R2 is not provided because it was very close to zero in all cases. Individual sector LR example plots for the Mo/Mo – 26 kVp settings are shown in Fig. 3 (glandular) and Fig. 4 (fatty). Plots for a given tissue type are shown in the same figure for intersystem comparisons. In these plots, measured points from the FL location are denoted by diamonds and asterisks for the KS measured points. Fitted lines from the regression analysis are solid for the FL plots and dashed for KS plots. Regression parameters are listed in Table 2. Regression parameter distribution quantities for the other Mo/Rh and Rh/Rh settings are shown in Table 3. The corresponding individual sector plot examples for these other settings are not shown because they were similar to the prior examples. For all distribution examples, variation within given location is relatively small, and the effective attenuation coefficients are in close agreement between the locations. Moreover, the linear model approximation holds when assessing the R2 statistic for all situations. To show the intersystem agreement, a paired t-test was applied by pairing all of the 10 average attenuation coefficient quantities for the FL location with the corresponding values from the KS location (second column from Table 1 and Table 3), which gave p = 0.56 (linear correlation of = 0.97). Following the same procedure for the 10 average intercept quantities (using the fifth columns from the same tables), gave p = 0.101 (linear correlation = 0.99). Using the p value as a gauge, effective attenuation coefficients as an ensemble are similar across the two systems. Intersystem intercept comparisons indicate a weak system dependence.

Bottom Line: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared.Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically.Calibration data collection can be reduced by applying a simple normalization technique.

View Article: PubMed Central - HTML - PubMed

Affiliation: Cancer Prevention & Control Department, Moffitt Cancer Center, Tampa, Florida, USA. john.heine@moffitt.org

ABSTRACT

Background: Breast density is a significant breast cancer risk factor. Currently, there is no standard method for measuring this important factor. Work presented here represents an essential component of an ongoing project that seeks to determine the appropriate method for calibrating (standardizing) mammography image data to account for the x-ray image acquisition influences. Longer term goals of this project are to make accurate breast density measurements in support of risk studies.

Methods: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared. Normalization methods were studied to assess the possibility of reducing the amount of calibration data collection. The percent glandular calibration map functional form was investigated. Spatial variations in the calibration data were used to assess the uncertainty in the calibration application by applying error propagation analyses.

Results: Logarithmic response curves are well approximated as linear. Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically. Calibration data collection can be reduced by applying a simple normalization technique. The calibration map is well approximated as linear. Intrasystem calibration variation was on the order of four percent, which was approximately half of the intersystem variation.

Conclusion: FFDM systems provide a quantitative output, and the calibration quantities presented here may be used for data acquired on similar FFDM systems.