Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications.

Heine JJ, Thomas JA - Biomed Eng Online (2008)

Bottom Line: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared.Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically.Calibration data collection can be reduced by applying a simple normalization technique.

View Article: PubMed Central - HTML - PubMed

Affiliation: Cancer Prevention & Control Department, Moffitt Cancer Center, Tampa, Florida, USA. john.heine@moffitt.org

ABSTRACT

Background: Breast density is a significant breast cancer risk factor. Currently, there is no standard method for measuring this important factor. Work presented here represents an essential component of an ongoing project that seeks to determine the appropriate method for calibrating (standardizing) mammography image data to account for the x-ray image acquisition influences. Longer term goals of this project are to make accurate breast density measurements in support of risk studies.

Methods: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared. Normalization methods were studied to assess the possibility of reducing the amount of calibration data collection. The percent glandular calibration map functional form was investigated. Spatial variations in the calibration data were used to assess the uncertainty in the calibration application by applying error propagation analyses.

Results: Logarithmic response curves are well approximated as linear. Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically. Calibration data collection can be reduced by applying a simple normalization technique. The calibration map is well approximated as linear. Intrasystem calibration variation was on the order of four percent, which was approximately half of the intersystem variation.

Conclusion: FFDM systems provide a quantitative output, and the calibration quantities presented here may be used for data acquired on similar FFDM systems.

Show MeSH

Related in: MedlinePlus

Rectangular phantom used in the Florida location. Labeled rectangular region is the field of view defined in the manuscript. The phantom nearly covers the entire detector.
© Copyright Policy - open-access
Related In: Results  -  Collection

License
getmorefigures.php?uid=PMC2365951&req=5

Figure 1: Rectangular phantom used in the Florida location. Labeled rectangular region is the field of view defined in the manuscript. The phantom nearly covers the entire detector.

Mentions: Phantom imaging was performed at two locations with GE Senographe 2000D FFDM systems that were manufactured about the same time (spring-summer 2000). Both systems are used for clinical breast screening. The systems are located at the Moffitt Cancer Center, Tampa, Florida, and the via Christy Regional Medical Center, Wichita, Kansas, which will be referenced as the FL and KS locations, respectively. Senographe 2000D systems produces both raw data (x-ray exposure representation) and processed data with 100 micron digital resolution and 14 bit dynamic pixel range. Raw data was used for this work. This system has three target\filter combinations (described below). The system detector is described elsewhere [13,15]. Phantoms used for this study were obtained from Computerized Imaging References Systems (CIRS, Norfolk VA). The quality of these phantoms and their close similarity with the x-ray interaction characteristics of breast tissue within the realm of mammography x-ray spectra are discussed in previous work [8] and related references therein. Phantoms for both locations have different geometries but were manufactured with the same material specifications: 100% glandular breast tissue equivalent and 100% fatty breast tissue equivalent. Florida phantoms are 18 cm × 24 cm rectangular slabs with either one or two cm heights. Kansas phantoms have a semicircular breast-like geometry with one cm heights. Example phantom images for the FL and KS locations are shown in Fig. 1 and Fig. 2, respectively. Kansas phantoms are 19 cm in length along the left hand margin in the y direction (from top to bottom), and the distance from left hand border along the x direction to the furthest distance to the semicircular perimeter is 13 cm.


Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications.

Heine JJ, Thomas JA - Biomed Eng Online (2008)

Rectangular phantom used in the Florida location. Labeled rectangular region is the field of view defined in the manuscript. The phantom nearly covers the entire detector.
© Copyright Policy - open-access
Related In: Results  -  Collection

License
Show All Figures
getmorefigures.php?uid=PMC2365951&req=5

Figure 1: Rectangular phantom used in the Florida location. Labeled rectangular region is the field of view defined in the manuscript. The phantom nearly covers the entire detector.
Mentions: Phantom imaging was performed at two locations with GE Senographe 2000D FFDM systems that were manufactured about the same time (spring-summer 2000). Both systems are used for clinical breast screening. The systems are located at the Moffitt Cancer Center, Tampa, Florida, and the via Christy Regional Medical Center, Wichita, Kansas, which will be referenced as the FL and KS locations, respectively. Senographe 2000D systems produces both raw data (x-ray exposure representation) and processed data with 100 micron digital resolution and 14 bit dynamic pixel range. Raw data was used for this work. This system has three target\filter combinations (described below). The system detector is described elsewhere [13,15]. Phantoms used for this study were obtained from Computerized Imaging References Systems (CIRS, Norfolk VA). The quality of these phantoms and their close similarity with the x-ray interaction characteristics of breast tissue within the realm of mammography x-ray spectra are discussed in previous work [8] and related references therein. Phantoms for both locations have different geometries but were manufactured with the same material specifications: 100% glandular breast tissue equivalent and 100% fatty breast tissue equivalent. Florida phantoms are 18 cm × 24 cm rectangular slabs with either one or two cm heights. Kansas phantoms have a semicircular breast-like geometry with one cm heights. Example phantom images for the FL and KS locations are shown in Fig. 1 and Fig. 2, respectively. Kansas phantoms are 19 cm in length along the left hand margin in the y direction (from top to bottom), and the distance from left hand border along the x direction to the furthest distance to the semicircular perimeter is 13 cm.

Bottom Line: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared.Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically.Calibration data collection can be reduced by applying a simple normalization technique.

View Article: PubMed Central - HTML - PubMed

Affiliation: Cancer Prevention & Control Department, Moffitt Cancer Center, Tampa, Florida, USA. john.heine@moffitt.org

ABSTRACT

Background: Breast density is a significant breast cancer risk factor. Currently, there is no standard method for measuring this important factor. Work presented here represents an essential component of an ongoing project that seeks to determine the appropriate method for calibrating (standardizing) mammography image data to account for the x-ray image acquisition influences. Longer term goals of this project are to make accurate breast density measurements in support of risk studies.

Methods: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared. Normalization methods were studied to assess the possibility of reducing the amount of calibration data collection. The percent glandular calibration map functional form was investigated. Spatial variations in the calibration data were used to assess the uncertainty in the calibration application by applying error propagation analyses.

Results: Logarithmic response curves are well approximated as linear. Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically. Calibration data collection can be reduced by applying a simple normalization technique. The calibration map is well approximated as linear. Intrasystem calibration variation was on the order of four percent, which was approximately half of the intersystem variation.

Conclusion: FFDM systems provide a quantitative output, and the calibration quantities presented here may be used for data acquired on similar FFDM systems.