Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension.

Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG - Vasc Health Risk Manag (2007)

Bottom Line: The number of differences in each category was compared with the number required by the International Protocol.For all devices, readings differing by less than 5, 10, and 15 mmHg for SBP and DBP values fulfill the recommendation criteria of the International Protocol as well as the individual analysis.Omron M1 Plus (HEM-401 1C-E), Omron M6 Comfort (HEM 7000-E), Spengler KP7500 D, and Microlife BP A100 Plus devices fulfilled the validation recommendations of the International Protocol.

View Article: PubMed Central - PubMed

Affiliation: The CardioVascular Institute, Paris, France.

ABSTRACT

Background: Four electronic devices for self-measurement of brachial blood pressure (BP): the Omron M1 Plus, the Omron M6 Comfort, the Spengler KP7500 D, and the Microlife BP A100 Plus, were evaluated in four separate studies according to the International Protocol of the European Society of Hypertension (ESH).

Design: The International Validation Protocol is divided into 2 phases: the first phase is performed on 15 selected subjects (45 pairs of BP measurements); if the device passes this phase, 18 supplementary subjects are included (54 pairs of BP measurements) making a total number of 33 subjects (99 pairs of BP measurements) on which the final validation is performed.

Methods: The same methodology recommended by the ESH protocol was applied for the 4 studies. In each study and for each subject, 4 BP measurements were performed simultaneously by 2 trained observers using mercury sphygmomanometers alternately with 3 measurements by the tested device. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. The 99 pairs of BP differences were classified into 3 categories (< or =5, 10, < or =15 mmHg). The number of differences in each category was compared with the number required by the International Protocol. An individual analysis was then done to determine for each subject the number of comparisons < or =5 mmHg. At least 22 of the 33 subjects should have 2 of their 3 comparisons < or =5 mmHg.

Results: All 4 tested devices passed the first and the second phase of the validation process. The average differences between the device and mercury sphygmomanometer readings were -1.4 +/- 5.5 and -0.4 +/- 4.8 mmHg for SBP and DBP respectively for the Omron M1 Plus device, -2.1 +/- 7.4 and 0.1 +/- 4.9 mmHg for SBP and DBP respectively for the Omron M6 Comfort device, -1.4 +/- 8.6 and -0.1 +/- 3.5 mmHg for SBP and DBP respectively for the Spengler KP7500 D device, and 1.6 +/- 4.2 mmHg and 0.54 +/- 2.8 mmHg for SBP and DBP respectively for the Microlife BP A100 Plus device. For all devices, readings differing by less than 5, 10, and 15 mmHg for SBP and DBP values fulfill the recommendation criteria of the International Protocol as well as the individual analysis.

Conclusions: Omron M1 Plus (HEM-401 1C-E), Omron M6 Comfort (HEM 7000-E), Spengler KP7500 D, and Microlife BP A100 Plus devices fulfilled the validation recommendations of the International Protocol.

Show MeSH

Related in: MedlinePlus

Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.
© Copyright Policy
Related In: Results  -  Collection


getmorefigures.php?uid=PMC2291343&req=5

fig2: Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.

Mentions: The difference between the device readings and the mean BP of the device and the two observers for all 99 points for SBP and DBP are shown in Figure 1a and Figure 2a respectively.


Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension.

Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG - Vasc Health Risk Manag (2007)

Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.
© Copyright Policy
Related In: Results  -  Collection

Show All Figures
getmorefigures.php?uid=PMC2291343&req=5

fig2: Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.
Mentions: The difference between the device readings and the mean BP of the device and the two observers for all 99 points for SBP and DBP are shown in Figure 1a and Figure 2a respectively.

Bottom Line: The number of differences in each category was compared with the number required by the International Protocol.For all devices, readings differing by less than 5, 10, and 15 mmHg for SBP and DBP values fulfill the recommendation criteria of the International Protocol as well as the individual analysis.Omron M1 Plus (HEM-401 1C-E), Omron M6 Comfort (HEM 7000-E), Spengler KP7500 D, and Microlife BP A100 Plus devices fulfilled the validation recommendations of the International Protocol.

View Article: PubMed Central - PubMed

Affiliation: The CardioVascular Institute, Paris, France.

ABSTRACT

Background: Four electronic devices for self-measurement of brachial blood pressure (BP): the Omron M1 Plus, the Omron M6 Comfort, the Spengler KP7500 D, and the Microlife BP A100 Plus, were evaluated in four separate studies according to the International Protocol of the European Society of Hypertension (ESH).

Design: The International Validation Protocol is divided into 2 phases: the first phase is performed on 15 selected subjects (45 pairs of BP measurements); if the device passes this phase, 18 supplementary subjects are included (54 pairs of BP measurements) making a total number of 33 subjects (99 pairs of BP measurements) on which the final validation is performed.

Methods: The same methodology recommended by the ESH protocol was applied for the 4 studies. In each study and for each subject, 4 BP measurements were performed simultaneously by 2 trained observers using mercury sphygmomanometers alternately with 3 measurements by the tested device. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. The 99 pairs of BP differences were classified into 3 categories (< or =5, 10, < or =15 mmHg). The number of differences in each category was compared with the number required by the International Protocol. An individual analysis was then done to determine for each subject the number of comparisons < or =5 mmHg. At least 22 of the 33 subjects should have 2 of their 3 comparisons < or =5 mmHg.

Results: All 4 tested devices passed the first and the second phase of the validation process. The average differences between the device and mercury sphygmomanometer readings were -1.4 +/- 5.5 and -0.4 +/- 4.8 mmHg for SBP and DBP respectively for the Omron M1 Plus device, -2.1 +/- 7.4 and 0.1 +/- 4.9 mmHg for SBP and DBP respectively for the Omron M6 Comfort device, -1.4 +/- 8.6 and -0.1 +/- 3.5 mmHg for SBP and DBP respectively for the Spengler KP7500 D device, and 1.6 +/- 4.2 mmHg and 0.54 +/- 2.8 mmHg for SBP and DBP respectively for the Microlife BP A100 Plus device. For all devices, readings differing by less than 5, 10, and 15 mmHg for SBP and DBP values fulfill the recommendation criteria of the International Protocol as well as the individual analysis.

Conclusions: Omron M1 Plus (HEM-401 1C-E), Omron M6 Comfort (HEM 7000-E), Spengler KP7500 D, and Microlife BP A100 Plus devices fulfilled the validation recommendations of the International Protocol.